Error: 105 PIC Motor Error indicating a motor failure flag that can cause infusion stoppage, inability to start infusion, or inconsistent pumping performance.
Step-by-Step Troubleshooting
1. Ensure Patient Safety First
- Remove the pump from patient use immediately.
- Transfer infusion to a backup pump.
2. Verify the Reported Error and Behavior
- Power on the pump and confirm Error: 105 PIC Motor Error is displayed.
- Attempt to start an infusion if safe to test.
- Observe for failure to start, immediate alarms, or intermittent stopping.
3. Check Power Source and Battery Condition
- Ensure the pump is connected to AC power.
- Verify battery charge level and charging indicators.
4. Inspect for Mechanical Obstruction
- Open the pump door and inspect the pumping channel.
- Check for kinked or improperly loaded tubing.
- Look for debris or residue inside the channel.
- Reload tubing using a known-good administration set.
5. Check Door Closure and Latch Integrity
- Close the pump door firmly and ensure it latches completely.
- Inspect for misalignment or damage to the latch.
- Verify the door is fully seated with no gaps.
6. Power Cycle the Pump
- Turn the pump off completely.
- Disconnect from AC power briefly, then reconnect and power on.
7. Isolate Device vs Setup
- Test with a different IV tubing set.
- Try a different power outlet.
- Compare with another SIGMA Spectrum pump if available.
If the Problem Persists
If Error: 105 PIC Motor Error continues after all checks, external causes have been ruled out and the motor failure flag indicates an internal fault.
- Remove device from service
- Label Out of Service
- Send for repair or bench evaluation
Clinical Use Tip
Never troubleshoot a pump on an active patient. Always move the patient to a backup device before evaluating motor-related errors.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Pump displaying Error: 105 PIC Motor Error and unable to start infusion."
Cause
What was observed during troubleshooting.
Example:
"Internal motor failure flag present; external causes ruled out."
Resolution
What action was taken.
Example:
"Removed from service and sent for motor/control evaluation."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
Motor-related errors directly impact infusion accuracy and patient safety. Always rule out external mechanical and setup causes first, then escalate promptly when internal faults are indicated.
That is successful troubleshooting.