What This Guide Helps With

Error 301 US upper avg high indicating abnormal ultrasonic sensor readings, often causing infusion interruption or false fluid detection issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check IV Set Installation and Compatibility

Inspect for Fluid Contamination in Pump Channel

Check for Air Bubbles or Improper Priming

Swap IV Set with Known Good Set

Power Cycle the Pump

Evaluate for Environmental Factors

If the Problem Persists

If Error 301 continues after all external checks, external causes have been ruled out and ultrasonic sensor readings remain out of range.

Clinical Use Tip

Never troubleshoot fluid detection or sensor-related errors while connected to a patient. Always transfer to a backup device first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump displaying Error 301 US upper avg high and not allowing infusion start"

Cause

What was observed during troubleshooting.

Example:
"Ruled out IV set misload and contamination; error persisted with known-good tubing, indicating likely ultrasonic sensor fault"

Resolution

What action was taken.

Example:
"Removed pump from service and sent for repair"

Helpful Details to Include (If Known)

Final Thought

Ultrasonic sensor errors are often caused by external setup or contamination issues. Always rule out tubing, priming, and cleanliness before escalating to internal failure. Proper troubleshooting ensures both safety and efficiency.

That is successful troubleshooting.