Baxter SIGMA Spectrum

Error: 306 DS rdng out of range (high)

On this page

Asset Type

Infusion Pump

Manufacturer

Baxter

Model

SIGMA Spectrum

What This Guide Helps With

Error 306 indicates a downstream (DS) reading out of range (high), which can stop infusion due to abnormal pressure conditions, occlusions, or sensor-related issues affecting safe fluid delivery.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check IV Set Installation and Compatibility

Inspect for Downstream Occlusion or Restriction

Check for Fluid or Contamination in Pump Channel

Replace the IV Set with a Known-Good Set

Restart the Pump (Power Cycle)

Attempt Operation Without Patient Connection (Test Setup)

If the Problem Persists

If Error 306 continues after all external checks:

Clinical Use Tip

Never troubleshoot pressure-related alarms while connected to a patient. Always transfer infusion to a backup device before continuing diagnostics.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump displaying Error 306 DS rdng out of range (high); infusion would not start."

Cause

What was observed during troubleshooting.

Example:
"High downstream pressure detected. External causes ruled out; suspected internal sensor fault."

Resolution

What action was taken.

Example:
"Replaced IV set and verified tubing path. Removed pump from service after error persisted."

Helpful Details to Include (If Known)

Final Thought

Error 306 indicates a downstream pressure or sensing issue. Always rule out occlusions and setup problems first before concluding a device fault. Proper troubleshooting ensures safe operation and efficient escalation when needed.

That is successful troubleshooting.