Baxter SIGMA Spectrum

Error: 314 DS no-tube elevated

On this page

Asset Type

Infusion Pump

Manufacturer

Baxter

Model

SIGMA Spectrum

What This Guide Helps With

Error 314 indicates the downstream (DS) pressure sensor is reading high even when no tubing is installed. This prevents the pump from passing self-check and safely operating due to unreliable occlusion detection.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Inspect Pump Channel for Obstruction or Contamination

Check for Mechanical Obstruction or Misalignment

Perform Power Cycle

Test Without Tubing Installed

Load Known-Good Tubing Set

If the Problem Persists

If Error 314 continues after cleaning, inspection, and testing with no tubing and known-good tubing, external causes have been ruled out.

Clinical Use Tip

Never troubleshoot a pump on an active patient. Always transition the patient to a backup infusion device before evaluating alarms or errors.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump displaying Error 314 DS no-tube elevated during startup; unable to initiate infusion."

Cause

What was observed during troubleshooting.

Example:
"Downstream pressure sensor reading high with no tubing present; no external obstruction or contamination found."

Resolution

What action was taken.

Example:
"Cleaned pump channel and verified proper operation. Error persisted; removed from service and sent for repair."

Helpful Details to Include (If Known)

Final Thought

Pressure sensor errors directly impact infusion safety and occlusion detection. Always rule out contamination and setup issues first, then escalate promptly when abnormal readings persist.

That is successful troubleshooting.