Asset Type

Infusion Pump

Manufacturer

Baxter

Model

SIGMA Spectrum

What This Guide Helps With

Error 322 Link Switch Error (low) indicates the lower link switch is not activating. This prevents proper cassette detection and can stop the pump from operating safely due to incorrect door or latch position sensing.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Inspect Door and Latch Mechanism

Check for Obstructions or Debris

Verify Cassette Installation

Test with Another Cassette

Observe Mechanical Linkage Movement

Perform Power Cycle

If the Problem Persists

If all external checks, cleaning, and cassette verification steps have been completed and the error remains, the lower link switch or internal linkage is likely faulty.

Clinical Use Tip

Never troubleshoot on an active patient. Always transfer the patient to a backup infusion pump before beginning troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump displaying Error 322 Link Switch Error (low) when loading cassette; infusion unable to start."

Cause

What was observed during troubleshooting.

Example:
"Lower link switch not activating. Verified door function, cassette seating, and no obstruction. Issue consistent with internal switch or linkage failure."

Resolution

What action was taken.

Example:
"Cleaned cassette area, verified proper cassette installation, tested with known-good cassette, and power cycled unit. Error persisted. Removed from service and sent for repair."

Helpful Details to Include (If Known)

Final Thought

Link switch errors indicate the pump cannot verify safe cassette engagement. Always perform simple mechanical checks first, but escalate quickly when the issue is consistent. Reliable detection systems are critical for safe infusion therapy.

That is successful troubleshooting.