Baxter SIGMA Spectrum

Error: 347 Check Battery

On this page

Asset Type

Infusion Pump

Manufacturer

Baxter

Model

SIGMA Spectrum

What This Guide Helps With

Error 347 Check Battery indicating defective battery causing power instability, failure to hold charge, or unexpected shutdown during operation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check AC Power Source and Power Adapter

Inspect Battery Installation

Check Charging Behavior

Attempt Operation on Battery Only

Swap with Known Good Battery

If the Problem Persists

If Error 347 remains after battery replacement and external checks:

Clinical Use Tip

Never troubleshoot on an active patient. Always move the patient to a backup device before evaluating battery or power-related issues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump displaying Error 347 Check Battery. Unit shuts down when unplugged."

Cause

What was observed during troubleshooting.

Example:
"Battery failed to hold charge and error persisted until replaced. Verified with known good battery."

Resolution

What action was taken.

Example:
"Replaced defective battery and verified proper charging and operation. Returned unit to service."

Helpful Details to Include (If Known)

Final Thought

Battery-related errors directly impact patient safety due to the risk of therapy interruption. Always verify reliable power operation and escalate promptly when internal faults are suspected.

That is successful troubleshooting.