Baxter SIGMA Spectrum Infusion Pump

Occlusion Alarm / Will Not Start Infusion

On this page

Asset Type

Infusion Pump

Manufacturer

Baxter

Model

SIGMA Spectrum

What This Guide Helps With

This guide provides a structured Clinical Engineering approach when a Baxter SIGMA Spectrum infusion pump will not start an infusion due to an occlusion alarm.

The focus is on logical, external, and easily verifiable causes before assuming internal pump failure. These steps prioritize patient safety, proper escalation, and clear documentation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Inspect the IV Set and Tubing Path

Check the Patient-Side of the Line

Inspect the Cassette and Pump Door

Confirm Fluid Source and Prime Status

Review Alarm Details on Display

Perform a Controlled Reset

If the Problem Persists

If the occlusion alarm continues after completing all external checks, common causes have been ruled out. At this point:

Knowing when to stop bedside troubleshooting is proper Clinical Engineering practice.

Clinical Use Tip

Never attempt extended troubleshooting on an active patient receiving critical medications.

Always transition the patient to a functioning infusion device first. Patient care continuity takes priority over equipment evaluation.

Work Order Documentation (CCR Method)

Complaint

What was reported by the clinical staff.

Example:
“Pump will not start infusion. Occlusion alarm displayed.”

Cause

What was observed during troubleshooting.

Example:
“Tubing inspected and free of kinks. Clamps open. Cassette reseated. Fluid present and line primed. Alarm persists after reload. Suspect internal pressure sensing issue.”

Resolution

What action was taken.

Example:
“Removed pump from service, labeled Out of Service, sent to CE shop for evaluation.”
or
“Bench tested; unable to clear occlusion alarm. Forwarded for repair.”

Helpful Details to Include (If Known)

Final Thought

An occlusion alarm is often caused by simple external flow restriction. Systematically checking tubing, clamps, cassette seating, and fluid source prevents unnecessary repairs and protects patient safety.

Clear documentation, logical troubleshooting, and proper escalation reflect strong Clinical Engineering practice.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide