BD Alaris Infusion Pump

Air-in-Line Alarms

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

This guide addresses repeated or unexpected air-in-line alarms on a BD Alaris infusion pump. It focuses on simple, external, and easily verifiable causes first—such as IV bag setup, tubing priming, cassette loading, and air sensor cleanliness—before assuming internal failure.

Air-in-line alarms are often setup-related. Logical troubleshooting can resolve most issues without unnecessary repair escalation.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Inspect the IV Bag or Fluid Container

3. Inspect the Tubing from Bag to Pump

4. Check the Pump Segment in the Mechanism

5. Inspect the Air Sensor Area

6. Replace the Administration Set

7. Test the Pump Off-Patient

If the Problem Persists

If all external causes have been ruled out and the air-in-line alarm continues during controlled testing, the likely causes may include:

At this point:

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never attempt to “clear alarms repeatedly” on an active patient without verifying the cause. Air-in-line alarms exist to prevent air embolism risk. Always ensure patient therapy is safely transferred before extended troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Nurse reports repeated air-in-line alarms during infusion despite full IV bag.”

Cause

What was observed during troubleshooting.

Example:
“Visible air bubbles noted in tubing due to incomplete priming. Tubing replaced and properly primed. No further alarms observed during functional test.”

Resolution

What action was taken.

Example:
“Replaced administration set, re-primed line, verified proper seating in pump channel. Performed off-patient functional test. Pump returned to service.”

“Removed from service. Tagged Out of Service and sent to shop for air sensor evaluation.”

Helpful Details to Include in Documentation

Final Thought

Air-in-line alarms are frequently caused by setup or tubing issues—not pump failure. A structured approach protects the patient, avoids unnecessary repairs, and supports reliable documentation. When external causes are eliminated and alarms persist, escalation is the correct and professional decision.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide