Error: 120.4310 Low voltage on 24V supply line

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

This guide helps troubleshoot Error 120.4310 Low voltage on 24V supply line on the BD Alaris Infusion Pump. This condition indicates insufficient internal power supply voltage, which can cause pump instability, charging failure, module communication issues, or complete loss of operation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check AC Power Source (Main PCU)

Inspect Power Cord and Inlet

Check Battery Status

Remove All Attached Modules

Inspect Module Connections and Contacts

Inspect for Physical Damage or Fluid Intrusion

Power Cycle the System

If the Problem Persists

If Error 120.4310 continues after all external checks:

Clinical Use Tip

Never troubleshoot power-related errors on an active patient. Always transfer the patient to a backup device before performing any checks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"BD Alaris PCU displaying Error 120.4310 with intermittent shutdowns during infusion."

Cause

What was observed during troubleshooting.

Example:
"Verified AC power, cord, battery, and modules. Error persisted, indicating internal 24V supply failure."

Resolution

What action was taken.

Example:
"Removed PCU from service and sent for repair and evaluation."

Helpful Details to Include (If Known)

Final Thought

Low voltage errors in the Alaris system can originate from external power issues, battery degradation, or module loading conditions. Always rule out these external factors before determining an internal failure requiring repair.

That is successful troubleshooting.