BD Alaris

Error: 120.4520 Battery did not respond appropriately to precharge current

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

Error 120.4520 indicates a failure in the battery precharge process, preventing the pump from recognizing or charging the battery. This results in loss of battery backup capability, which can interrupt therapy during transport or power loss.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Check AC Power and Charging Behavior

4. Reseat the Battery

5. Inspect Battery and Contacts

6. Swap with a Known Good Battery

7. Power Cycle the Device

If the Problem Persists

If Error 120.4520 continues after reseating the battery, inspecting contacts, and replacing with a known good battery, this indicates failure of the internal power management or battery precharge circuitry.

Clinical Use Tip

Never troubleshoot battery or power issues on an active patient. Always move the patient to a backup device before continuing troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"BD Alaris pump displaying Error 120.4520; battery not recognized or not charging."

Cause

What was observed during troubleshooting.

Example:
"Battery failed precharge response. Error persisted after reseating and replacement with known good battery, indicating internal power circuit issue."

Resolution

What action was taken.

Example:
"Removed from service, labeled out of service, and sent for repair of battery charging circuitry."

Helpful Details to Include (If Known)

Final Thought

Battery-related errors directly affect safe infusion operation. Always rule out external causes first, but escalate quickly when internal circuitry failure is suspected.

That is successful troubleshooting.