BD Alaris

Error: 222.4010 High infusion rate detected

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

This guide addresses Error: 222.4010 High infusion rate detected, which occurs when the pump identifies fluid delivery exceeding programmed or safe limits. The pump stops infusion to prevent potential over-infusion, medication overdose, or fluid overload.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Check Infusion Programming

4. Verify Drug Library / Guardrails

5. Inspect Administration Set and Pump Loading

6. Check for Free Flow or Gravity-Assisted Flow

7. Check for Occlusion Conditions

8. Power Cycle the Pump Module

9. Test with Known Good Setup

If the Problem Persists

If Error: 222.4010 continues after verifying programming and setup, external causes have been ruled out and the issue likely indicates an internal flow sensing or safety processor fault.

Clinical Use Tip

Never troubleshoot or resume infusion on an active patient. Always move the patient to a backup device before continuing investigation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"BD Alaris pump displaying Error: 222.4010 and stopping infusion."

Cause

What was observed during troubleshooting.

Example:
"Verified programming and tubing setup; no external issues found. Suspected internal flow sensing fault."

Resolution

What action was taken.

Example:
"Replaced administration set and confirmed settings; error persisted. Removed pump from service and sent for repair."

Helpful Details to Include (If Known)

Final Thought

High infusion rate detection is a critical safety feature. Always verify setup thoroughly, but escalate quickly if the issue persists to prevent patient harm.

That is successful troubleshooting.