BD Alaris

Error: 222.4180 Improper door and/or flow stop state detected during infusion

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

Error 222.4180 indicating a mismatch between door position and flow stop state, preventing infusion due to unsafe or unverified tubing conditions.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Inspect Pump Module Door

4. Check Roller Clamp / Flow Stop on IV Set

5. Reseat the IV Set Properly

6. Try a New IV Set

7. Inspect Door and Latch Area (External Only)

8. Power Cycle the Module

9. Perform Test Infusion (Off Patient)

If the Problem Persists

If external causes have been ruled out and Error 222.4180 continues:

Clinical Use Tip

Never troubleshoot on an active patient. Always transfer therapy to a backup device before investigating pump errors.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump displaying Error 222.4180 and infusion would not start."

Cause

What was observed during troubleshooting.

Example:
"Improper door/flow stop state detected. Verified proper tubing load and clamp position; issue persisted."

Resolution

What action was taken.

Example:
"Reloaded tubing and replaced IV set. Issue persisted, removed module from service for repair."

Helpful Details to Include (If Known)

Final Thought

Errors related to door and flow stop conditions are critical safety checks tied to anti-free-flow protection. Always verify setup and external factors first, but escalate quickly when safety sensing cannot be confirmed.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?