BD Alaris

Error: 222.4200 Motor activity detected when motor should be idle

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

Error 222.4200 indicating unintended motor movement while idle, risking uncontrolled infusion or pump shutdown.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Identify the Affected Module

Check Administration Set Installation

Inspect Pump Channel and Mechanism

Power Cycle the System

Check for Environmental or Handling Factors

Swap Administration Set

Perform Functional Test (Off Patient)

If the Problem Persists

External causes and module-swappable issues have been ruled out. This indicates a likely internal failure involving motor control circuitry, encoder feedback, or the safety processor.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device first before evaluating motor-related faults.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"BD Alaris infusion module displaying Error 222.4200 with motor activity while idle."

Cause

What was observed during troubleshooting.

Example:
"Unintended motor movement observed with no infusion running. Issue persisted after set replacement and module swap, indicating internal fault."

Resolution

What action was taken.

Example:
"Removed from service, labeled out of service, and sent for repair/vendor evaluation."

Helpful Details to Include (If Known)

Final Thought

Motor activity when idle is a serious safety concern. Once external factors and modular components are ruled out, escalation is the safest and correct path.

That is successful troubleshooting.