BD Alaris

Error: 241.4140 Patient-Side Pressure Sensor Voltage too low

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

This guide addresses Error 241.4140 indicating low patient-side pressure sensor voltage, which can lead to inaccurate downstream pressure monitoring, missed occlusions, or interruption of infusion therapy.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check Administration Set Compatibility and Installation

Inspect Patient-Side Tubing for Occlusion or Restriction

Inspect Pump Channel for Contamination or Damage

Replace with a Known-Good Administration Set

Power Cycle the Pump Module

Test Operation Off Patient

If the Problem Persists

If Error 241.4140 continues after all external checks:

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot pressure-related errors on an active patient. Always transfer the patient to a backup infusion device before continuing diagnostics.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"BD Alaris pump displaying Error 241.4140, infusion stopped during use."

Cause

What was observed during troubleshooting.

Example:
"Error persisted after replacing administration set and verifying tubing; no external obstruction found."

Resolution

What action was taken.

Example:
"Removed pump from service and sent for repair due to suspected internal pressure sensor failure."

Helpful Details to Include (If Known)

Final Thought

Patient-side pressure sensor errors directly affect infusion safety and occlusion detection. Always rule out external setup issues first, but escalate quickly once internal failure is suspected.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide