BD Alaris

Error: 241.4150 Patient-Side Pressure Sensor Voltage too high

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

This guide addresses Error 241.4150, indicating high voltage from the patient-side pressure sensor. This condition can lead to false downstream occlusion alarms, inaccurate pressure monitoring, or unexpected infusion interruption.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Check for Downstream Occlusion or Restriction

4. Verify IV Set Type and Proper Loading

5. Inspect Pump Channel and Pressure Interface

6. Power Cycle the Pump

7. Replace IV Set and Retest

8. Determine if Issue Persists

If the Problem Persists

If all external causes have been ruled out and the error continues, remove the device from service, label it as Out of Service, and send it for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot a pump with active pressure sensor errors on a patient. Always transition the patient to a backup device before beginning troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump displaying Error 241.4150 with repeated occlusion alarms and stopping infusion."

Cause

What was observed during troubleshooting.

Example:
"No downstream occlusion found, IV set properly loaded, channel clean. Error persisted, indicating likely pressure sensor fault."

Resolution

What action was taken.

Example:
"Removed pump from service, labeled out of service, and sent for repair."

Helpful Details to Include (If Known)

Final Thought

Always prioritize external causes like occlusions and setup issues before suspecting internal faults. Persistent sensor voltage errors indicate deeper system issues that require escalation and proper documentation.

That is successful troubleshooting.