BD Alaris

Error: 242.4010 Fluid delivery mechanism turning too quickly (infusion rate high)

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

Error 242.4010 indicates the fluid delivery mechanism is rotating faster than expected, suggesting a high infusion rate condition. This can lead to over-infusion, inaccurate delivery, or automatic pump shutdown to protect the patient.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Check Infusion Programming

4. Inspect IV Set Installation

5. Check for Free Flow or Low Resistance

6. Inspect Pump Channel Condition

7. Power Cycle and Retest

If the Problem Persists

If Error 242.4010 continues after all external checks:

Clinical Use Tip

Never attempt to troubleshoot or continue infusion on an active patient when high-rate errors occur. Always move the patient to a backup device first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"BD Alaris pump displaying Error 242.4010 with high infusion rate alarm during use."

Cause

What was observed during troubleshooting.

Example:
"Verified correct programming and IV set installation; error persisted indicating motor/encoder control issue."

Resolution

What action was taken.

Example:
"Removed module from service, labeled out of service, and sent for repair evaluation."

Helpful Details to Include (If Known)

Final Thought

High infusion rate errors indicate a loss of control over fluid delivery accuracy. After ruling out setup and external causes, escalate quickly to ensure patient safety and maintain device reliability.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide