BD Alaris

Error: 242.4030 Motor stall detected

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

Error 242.4030 motor stall causing infusion to stop or fail due to resistance, set misload, or pump drive mechanism issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check Administration Set Installation

Inspect for Kinks or Downstream Occlusion

Check Door Mechanism and Pump Channel

Replace Administration Set

Observe Pump Self-Test / Motor Behavior

If the Problem Persists

If the error continues after external causes such as tubing misload, occlusion, or set issues are ruled out:

Clinical Use Tip

Never troubleshoot an infusion pump on an active patient. Always transfer therapy to a verified working device first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump displayed Error 242.4030 and would not start infusion."

Cause

What was observed during troubleshooting.

Example:
"Motor stall condition persisted after verifying proper set installation and eliminating downstream occlusion."

Resolution

What action was taken.

Example:
"Replaced administration set and verified setup. Error persisted. Removed pump from service and sent for repair."

Helpful Details to Include (If Known)

Final Thought

Motor stall errors should always be approached by ruling out resistance and setup issues first. Once external causes are eliminated, escalation is the correct and safe decision.

That is successful troubleshooting.