BD Alaris

Error: 243.4070 Failed self test

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

Error 243.4070 indicating air-in-line subsystem self-test failure, preventing infusion due to inability to verify air detection safety function.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check for Set Loading and Channel Setup Issues

Inspect Air-in-Line Sensor Area

Power Cycle the Pump

Test with No Set Installed

Test with a Known-Good Set

If the Problem Persists

If Error 243.4070 continues after verifying setup, cleaning the sensor area, power cycling, and testing with and without a set, external causes have been ruled out.

Clinical Use Tip

Never troubleshoot on an active patient. Always move the patient to a backup device before attempting any troubleshooting steps.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"BD Alaris pump displaying Error 243.4070 during startup, preventing infusion"

Cause

What was observed during troubleshooting.

Example:
"Air-in-line subsystem failed internal self-test; error persisted after reseating set, cleaning sensor, and testing with known-good tubing"

Resolution

What action was taken.

Example:
"Removed pump from service, labeled out of service, and sent for repair"

Helpful Details to Include (If Known)

Final Thought

Self-test failures involving air detection are high-risk conditions. Proper troubleshooting focuses on ruling out setup and contamination issues quickly, then escalating appropriately to protect patient safety.

That is successful troubleshooting.