BD Alaris

Error: 244.3020 Clamp read CLOSED when it should be OPEN (unexpected state)

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

Error 244.3020 indicating a clamp state mismatch where the system detects the clamp as CLOSED when it should be OPEN. This prevents infusion from starting or continuing due to incorrect safety verification of tubing flow control.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check for Proper Administration Set Installation

Inspect Flow Stop / Safety Clamp Interaction

Check for Debris or Residue in Clamp Area

Verify Door Latch and Alignment

Swap Administration Set

Power Cycle the Module / PCU

If the Problem Persists

If all external causes have been ruled out and the error continues:

Clinical Use Tip

Never troubleshoot an infusion pump with clamp-related errors while connected to a patient. Always move the patient to a backup device before investigating the issue.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"BD Alaris module displaying Error 244.3020; infusion would not start due to clamp state error."

Cause

What was observed during troubleshooting.

Example:
"Clamp/flow stop state mismatch persisted after set replacement, cleaning, and door alignment checks; suspected sensor failure."

Resolution

What action was taken.

Example:
"Removed module from service, labeled Out of Service, and sent to biomed/vendor for repair."

Helpful Details to Include (If Known)

Final Thought

Clamp and flow stop mechanisms are critical safety features on infusion pumps. Always rule out setup and mechanical causes first, but escalate quickly when sensor feedback is unreliable.

That is successful troubleshooting.