BD Alaris

Error: 352.6640 Output is zero or outside acceptable range

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

This guide addresses Error 352.6640 indicating the force sensor output is zero or outside acceptable range. This condition prevents accurate pressure monitoring, which can lead to undetected occlusions or unsafe infusion delivery.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Check IV Set Installation

4. Inspect Pumping Channel and Sensor Area

5. Try a Known-Good IV Set

6. Power Cycle the Pump

7. Check for Door or Latch Issues

8. Perform Basic Functional Test

If the Problem Persists

If the error continues after verifying setup, cleaning, and replacing the IV set, external causes have been ruled out. This indicates a likely internal force sensor or circuitry failure.

Clinical Use Tip

Never troubleshoot infusion pumps on an active patient. Always transfer the patient to a backup device before addressing pressure or sensor-related issues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"BD Alaris pump displaying Error 352.6640; infusion would not start."

Cause

What was observed during troubleshooting.

Example:
"Force sensor output remained zero/out of range after verifying proper set installation and replacing IV set."

Resolution

What action was taken.

Example:
"Removed device from service and sent for repair due to suspected internal sensor failure."

Helpful Details to Include (If Known)

Final Thought

Force sensor errors directly impact occlusion detection and infusion safety. Always rule out setup and disposable issues first, but escalate promptly when the error persists.

That is successful troubleshooting.