BD Alaris Infusion Pump

Flow Rate Inaccurate or Infusion Not Delivering

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting situations where a BD Alaris infusion pump appears to deliver medication at an incorrect flow rate or does not deliver fluid as expected. Users may report that the infusion is running slower or faster than programmed, the fluid level is not decreasing, or the pump indicates it is infusing but little or no medication is reaching the patient.

Many of these issues are caused by IV set installation problems, flow restrictions, or configuration errors rather than an internal pump failure. This guide focuses on verifying those external causes first.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the reported issue

Confirm the programmed infusion settings

Inspect the IV set installation in the pump channel

Verify the correct IV set type is being used

Check for kinks, clamps, or restrictions in the IV line

Inspect the drip chamber and fluid source

Check for upstream or downstream occlusions

Confirm the pump door is fully closed and latched

Run a test infusion using a known-good IV set

Observe pump operation during the test infusion

If the Problem Persists

If the infusion pump continues to deliver inaccurate flow or fails to move fluid after verifying programming, tubing installation, and external restrictions, the issue may be related to an internal pumping mechanism, calibration drift, or sensor failure.

At this point the device should be:

Knowing when to stop troubleshooting prevents unnecessary risk and ensures the device receives proper evaluation.

Clinical Use Tip

Never troubleshoot an infusion pump while it is actively connected to a patient. If a flow accuracy issue is suspected, the patient should be transferred to another infusion pump before the device is evaluated. Maintaining uninterrupted and accurate medication delivery is critical for patient safety.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Nurse reported infusion running at 50 mL/hr but medication bag level not decreasing and patient not receiving expected volume.”

Cause

What was observed during troubleshooting.

Example:
“IV tubing found partially kinked near pump channel causing restricted flow.”

Resolution

What action was taken.

Example:
“Tubing repositioned and replaced with new compatible IV set. Pump tested and confirmed normal flow.”

Helpful Details to Include (If Known)

Final Thought

Accurate infusion delivery is critical to patient care, which makes systematic troubleshooting essential. Always begin with programming verification, tubing setup, and simple mechanical checks before suspecting internal failure. When external causes have been ruled out, removing the device from service and documenting findings ensures patient safety and supports effective repair.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide