BD Alaris Infusion Pump

Network Connectivity Failure (Wireless / Ethernet)

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting BD Alaris infusion pumps that fail to connect to the hospital network, either via wireless (Wi-Fi) or wired Ethernet. Symptoms may include the pump failing to communicate with the EMR, Guardrails Library updates not completing, or alarms indicating network errors. This guide emphasizes external, easily verifiable causes before considering internal network module issues.

Step-by-Step Troubleshooting

Verify Network Settings on Pump

Inspect Physical Connections (Ethernet Only)

Check Wi-Fi Signal and Coverage

Verify Network Availability

Restart Network Services on Pump

Check for IP Conflicts or DHCP Issues

Observe Network Indicators and Logs

If the Problem Persists

External network issues, cables, or settings have been ruled out.

Likely causes include internal network module failure or corrupted network configuration.

Clinical Use Tip

Never troubleshoot the network while the pump is actively infusing a patient.

Move the patient to another pump if necessary.

Ensure patient safety takes priority over connectivity troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump not communicating with EMR; wireless connection failing.”

Cause

What was observed during troubleshooting.

Example:
“Pump unable to connect; verified network settings, Wi-Fi coverage, and Ethernet cable; other devices on network functional.”

Resolution

What action was taken.

Example:
“Pump removed from service, labeled Out of Service, sent to Clinical Engineering for network module evaluation.”

Helpful Details to Include

Final Thought

Patient safety comes first; network connectivity is critical for EMR integration but should never compromise an active infusion. Logical, stepwise troubleshooting ensures external causes are resolved before escalating to internal repair. Accurate documentation in CCR format helps track patterns and supports cost avoidance when vendor intervention is required.

That is successful troubleshooting.