BD Alaris Infusion Pump

Occlusion Alarms

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

This guide addresses situations where the BD Alaris infusion pump triggers occlusion alarms. It focuses on easily verifiable, external causes—such as tubing kinks, connector issues, or pump setup errors—before assuming internal hardware failure. The goal is safe troubleshooting without disturbing a patient or the integrity of the infusion.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Check the Tubing Path

Verify Pump Cassette Seating

Check the Tubing Connections

Assess the Infusion Set

Examine the Pump and Pressure Sensor Area

Check Pump Settings and Alarms

If the Problem Persists

All external causes have been ruled out. The pump may have a faulty occlusion sensor or internal tubing mechanism.

Clinical Engineering should avoid further internal troubleshooting to prevent damage or patient risk.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump alarming occlusion despite no visible obstruction.”

Cause

What was observed during troubleshooting.

Example:
“Tubing kinked under bedrail and cassette slightly unseated.”

Resolution

What action was taken.

Example:
“Repositioned tubing, reseated cassette, verified alarm cleared, pump returned to service.”

Helpful Details to Include

Final Thought

Patient safety is paramount: troubleshooting must start with external, verifiable issues and never compromise active infusions. Proper documentation, logical assessment, and knowing when to escalate are the marks of effective Clinical Engineering.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide