BD Alaris Infusion Pump

Pump Reboots or Freezes During Operation

On this page

Asset Type

Infusion Pump

Manufacturer

BD

Model

Alaris

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting BD Alaris infusion pumps that unexpectedly reboot, freeze, or lock up during use. Symptoms may include the pump stopping mid-infusion, the touchscreen becoming unresponsive, alarms failing to trigger, or the device cycling power on its own. Most issues of this type can be related to power inconsistencies, software glitches, or peripheral connections rather than internal board-level failures.

Step-by-Step Troubleshooting

Verify Power Source

Why: Unstable power can cause unexpected reboots.

Inspect the Battery

Why: A failing battery can cause freezing or intermittent operation.

Examine External Cables and Accessories

Why: A faulty channel module or loose connection can disrupt pump operation.

Check for Environmental Factors

Why: Extreme temperatures can cause the microprocessor to reset.

Restart and Monitor

Why: Temporary software glitches can sometimes be resolved with a reboot.

Software / Guardrails Check

Why: Software or library corruption can lead to mid-operation freezes.

If the Problem Persists

Common external causes have been ruled out. The issue is likely internal (e.g., motherboard, memory module, or software corruption).

Knowing when to stop prevents unsafe troubleshooting and protects patient safety.

Clinical Use Tip

Do not troubleshoot on an active patient.

Move the patient to another functioning infusion pump.

Ensure the current pump is powered off and secured before any work.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump freezes mid-infusion or reboots unexpectedly during operation.”

Cause

What was observed during troubleshooting.

Example:
“External inspection revealed no cable issues, battery fully charged, power stable, but pump continues to reboot—likely internal software or hardware issue.”

Resolution

What action was taken.

Example:
“Device removed from service, labeled Out of Service, sent to vendor/biomed bench for evaluation.”

Helpful Details to Include

Final Thought

Consistently using a logical, stepwise troubleshooting approach ensures patient safety and reduces unnecessary repairs. Proper documentation and timely escalation maintain operational integrity and compliance standards.

That is successful troubleshooting.