On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoot synchronization loss between paired instruments or redundant systems caused by network, server availability, time, connectivity, configuration, or infrastructure problems.
Step-by-Step Troubleshooting
1. Protect Specimen Processing and Data Availability
Determine whether synchronization loss affects bottle status, results, worklists, alarms, or access to patient testing information.
Coordinate with laboratory staff before restarting servers, instruments, switches, or interfaces. Maintain an alternate verified workflow if data availability or result reporting is uncertain.
Expected outcome: Testing continuity is protected and no uncontrolled restart disrupts active blood culture processing.
2. Confirm Which Systems Are Out of Synchronization
Identify the affected components:
- One BACTEC FX instrument.
- Both instruments in a paired configuration.
- Primary server.
- Redundant or backup server.
- Workstation.
- Middleware or LIS interface.
- Network communication between components.
Determine when synchronization was last known to be normal.
Expected outcome: The synchronization problem is narrowed to a specific instrument, server, or communication path.
3. Check for Recent Infrastructure Changes
Ask whether the issue began after:
- Network maintenance.
- Switch replacement.
- VLAN or port change.
- Server reboot.
- Power interruption.
- Software update.
- Instrument relocation.
- Cable replacement.
- Planned downtime.
Do not assume the instrument is defective when the timing corresponds with an infrastructure event.
Expected outcome: A recent environmental or infrastructure change is either identified or ruled out.
4. Verify Power and Operating State of All Required Components
Confirm that required instruments, servers, workstations, and approved network equipment are powered and operating normally.
Look for:
- Unexpected shutdown.
- Boot screen.
- Application not running.
- Server unavailable message.
- UPS or power event.
- Repeated rebooting.
Expected outcome: All required components are powered, fully started, and available. If restoring a previously offline component restores synchronization, troubleshooting can stop after verification.
5. Inspect Physical Network Connections
Inspect accessible network connections for:
- Disconnected cables.
- Loose connectors.
- Damaged latches.
- Kinked or damaged patch cables.
- Connection to an incorrect wall jack or switch port after recent service.
- Missing link indication where normally expected.
Reseat accessible connections when permitted.
Expected outcome: Physical network connections are secure and normal link status is present. If reseating a loose connection restores synchronization, proceed to final verification.
6. Compare Network Availability Between the Paired Systems
Determine whether both instruments can communicate with the required server or network services through approved diagnostic methods.
Compare:
- One functioning instrument with the affected instrument.
- Network connection status.
- Server reachability.
- Whether other laboratory systems on the same network are affected.
Do not change IP addresses, VLANs, DNS settings, or other network configuration without coordination and authorization.
Expected outcome: The issue is identified as instrument-specific or infrastructure-wide.
7. Verify Server and Application Availability
Confirm that the required server and associated application services are available through normal supported methods.
If the redundant server is involved, determine whether:
- Both servers are online.
- The expected primary/secondary relationship is intact.
- One server recently rebooted.
- Application services are stopped or unavailable.
- Storage or operating-system warnings are present.
Do not modify server services, clustering, databases, or replication settings unless specifically authorized and qualified to do so.
Expected outcome: Required server resources are available, or the fault is appropriately transferred to the server or IT support team.
8. Check Date and Time Consistency
Verify that the paired instruments and servers show reasonable and consistent date and time information.
A significant time mismatch may interfere with event ordering, authentication, or synchronization depending on system architecture.
Do not manually alter system clocks in a validated environment without approved procedure.
Expected outcome: System time appears consistent. Any significant mismatch is escalated for authorized correction.
9. Review User-Accessible Communication Status
Use normal authorized system status screens to look for indications of:
- Disconnected server.
- Communication unavailable.
- Pending data.
- Offline partner.
- Failed interface.
- Unavailable redundant system.
Record the exact wording rather than clearing or repeatedly restarting the system.
Expected outcome: The communication failure is characterized without changing restricted configuration.
10. Coordinate an Approved Restart Only When Appropriate
If manufacturer documentation, laboratory workflow, and local IT policy permit a controlled restart, coordinate it with laboratory and IT personnel.
Restart only the affected component or system using the approved sequence. Do not arbitrarily reboot the entire paired environment during active testing.
Expected outcome: Required services reconnect and synchronization resumes. If synchronization remains stable and all pending data reconciles normally, troubleshooting can stop.
11. Verify End-to-End Synchronization
After correction, confirm that:
- Both instruments show the expected connection state.
- Required servers are available.
- New bottle or status information appears where expected.
- Results or events transfer appropriately.
- No persistent communication warning remains.
- The laboratory confirms normal workflow.
Expected outcome: Paired-system synchronization is restored and remains stable. Troubleshooting can stop and normal operation can resume.
12. Escalate Persistent Synchronization Loss
If cables, power, server availability, basic network connectivity, and approved configuration checks are normal but synchronization does not recover, stop external troubleshooting.
Expected outcome: The unresolved communication or replication problem is escalated without unauthorized changes to validated system configuration.
If the Problem Persists
Common external causes such as power loss, disconnected network cables, unavailable servers, recent infrastructure changes, time inconsistency, and basic communication availability have been ruled out.
The remaining problem may involve server replication, database synchronization, validated application services, instrument communication services, network configuration, software compatibility, or other service-level infrastructure.
The affected system should be:
- Removed from normal dependence when synchronized data integrity cannot be assured.
- Labeled Out of Service where appropriate.
- Evaluated by Clinical Engineering, IT, laboratory informatics, or the manufacturer according to system ownership.
- Evaluated using current manufacturer documentation and approved diagnostic tools.
- Reconfigured or repaired only by qualified personnel.
Before return to service, verify both instruments, the redundant server relationship, required interfaces, and expected data flow. Confirm that pending or previously unsynchronized records have been reconciled according to laboratory and manufacturer procedures.
Knowing when to stop external troubleshooting and involve IT, informatics, or manufacturer support is proper troubleshooting.
Clinical Use Tip
When paired blood culture systems lose synchronization, verify the complete data path before assuming results available on one instrument are visible everywhere they are clinically needed.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Laboratory staff reported that one BACTEC FX instrument stopped synchronizing bottle status with the paired system and server."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the affected instrument's network patch cable partially disconnected after recent equipment movement."
Resolution
What action was taken.
Example:
"Reseated the network connection and verified stable communication, synchronized bottle status, and normal data transfer between the instrument pair and server."
Helpful Details to Include (If Known)
- Instruments or servers affected
- Time synchronization was last known to work
- Recent network or server maintenance
- Power status of each component
- Network cable and link status
- Wall jack or switch port used
- Server/application availability
- Date and time consistency
- Exact communication message
- Whether other laboratory systems were affected
- Results before and after correction
- Confirmation of synchronized data
- Final device status
Final Thought
Protect data integrity and specimen workflow first, then check power, physical connections, server availability, network status, and authorized configuration before suspecting an internal software failure. Escalate persistent synchronization problems to the appropriate technical teams and document both the fault and verified restoration of the complete communication path.
That is successful troubleshooting.