BD Pyxis MedStation ES

Printer Will Not Print Labels or Receipts

On this page

Asset Type

Medication Dispensing Cabinet

Manufacturer

BD

Model

Pyxis MedStation ES

What This Guide Helps With

The cabinet printer does not produce usable output because of paper, loading, jams, power, connection, configuration, or printer hardware problems.

Step-by-Step Troubleshooting

1. Protect Clinical Workflow

Determine whether the missing printout affects medication identification, documentation, labeling, or another required clinical process.

Use the approved alternate workflow if printed output is required for safe medication use.

Expected outcome: Patient care continues without relying on a failed printer.

2. Confirm the Exact Printer Symptom

Determine whether the printer:

Expected outcome: The failure is classified as media, feed, print-quality, communication, or printer-power related.

3. Verify Paper or Label Supply

Inspect the accessible media supply.

Confirm:

Do not substitute unapproved labels or paper.

Expected outcome: The correct print media is installed and positioned properly.

4. Check for Accessible Jam or Obstruction

Inspect the normal media path for visible:

Remove only obstructions accessible through normal operator-accessible areas.

Expected outcome: The media path is clear.

If clearing an accessible jam restores normal printing, proceed to final verification and stop troubleshooting.

5. Inspect the Printer Externally

Check the printer housing, cover, latch, accessible cabling, and mounting for:

Expected outcome: The printer is physically intact and correctly closed or connected.

6. Verify Cabinet and Printer Power State

Confirm the cabinet is fully operational and look for normal printer indicators if externally visible.

A printer that is completely inactive may have a local power or connection issue.

Expected outcome: The printer has normal external signs of power or the lack of printer power is clearly identified.

7. Test a Known-Good Print Function

Using an approved non-clinical test or pharmacy-authorized print workflow, determine whether any output can be produced.

Avoid generating unnecessary patient or medication labels.

Expected outcome: The printer either produces a normal test output or consistently fails.

8. Evaluate Print Quality

If printing occurs, inspect for:

Do not perform internal printhead adjustment or calibration unless authorized by manufacturer procedure.

Expected outcome: Output is clearly readable and properly fed.

9. Verify Application and Configuration

If the printer is powered and mechanically functional but receives no jobs, coordinate with pharmacy informatics or authorized support to verify the intended printer and application configuration.

Do not change routing or print configuration without approval.

Expected outcome: The print job is correctly routed to the intended printer.

10. Perform Final Verification or Escalate

Verify a complete approved print job produces readable output without jamming.

If the printer remains unreliable after media, jam, power, external connection, and application checks, stop troubleshooting and escalate.

Expected outcome: Printing is reliable, or the cabinet/printer is removed from workflows requiring printed output.

If the Problem Persists

Common media, jam, external connection, and application-routing causes have been ruled out.

Possible remaining categories include printer mechanism, internal power, printhead, controller, internal cabling, or application-service problems.

The cabinet should be:

Return to service only after reliable print quality and correct application output have been verified.

Knowing when to stop before attempting internal printer repair is proper troubleshooting.

Clinical Use Tip

Unreadable or incomplete labels can be as unsafe as no label at all; verify print quality, not just that paper comes out.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pharmacy reported the Pyxis MedStation ES printer would feed but no usable receipt was produced."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the print media loaded incorrectly and folded in the accessible feed path."

Resolution

What action was taken.

Example:
"Removed the damaged media, reloaded approved media correctly, and verified a complete readable test print with normal feeding."

Helpful Details to Include (If Known)

Final Thought

Check media, jams, and external connections before suspecting printer hardware, verify that output is readable and correctly routed, and escalate when internal service is required.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide