Beckman Coulter DxC 700 AU

Probe, Pipettor, or Aspiration Error Occurs

On this page

Asset Type

Clinical Chemistry Analyzer

Manufacturer

Beckman Coulter

Model

DxC 700 AU

What This Guide Helps With

Use when aspiration, pipetting, or probe-related errors interrupt sample, reagent, calibration, or QC processing.

Step-by-Step Troubleshooting

1. Protect Patient Results and Stop Repeated Processing

Stop affected testing until reliable aspiration can be confirmed. Do not repeatedly process patient specimens through an analyzer producing recurring aspiration errors.

Expected outcome: Potentially incomplete or inaccurate pipetting does not affect released patient results.

2. Confirm When the Error Occurs

Determine whether the issue affects patient samples, reagents, calibrators, QC, one position, or all aspiration events. Record any displayed message.

Expected outcome: The problem is narrowed to a specific material, position, or broader pipetting function.

3. Inspect the Sample or Reagent Container

Check for insufficient volume, foam, bubbles, clotting, fibrin, debris, damaged containers, incorrect tube placement, or caps that could interfere with aspiration.

Expected outcome: The material is accessible and suitable for reliable aspiration.

4. Inspect Accessible Probe Areas Safely

With the analyzer in a safe nonmoving state, visually inspect accessible probe areas for visible residue, bending, obstruction, or abnormal positioning. Do not manually bend, align, or disassemble the probe.

Expected outcome: No obvious external obstruction or physical damage is present.

5. Verify Racks, Cups, and Holders

Confirm tubes, sample cups, reagent containers, and carriers are seated correctly and are appropriate for the analyzer workflow.

Expected outcome: Containers are positioned consistently in the expected aspiration location.

6. Use a Known-Good Sample or Control Setup

When permitted, test using an approved known-good container and material. If aspiration succeeds, compare the original specimen or container for clotting, bubbles, inadequate volume, or placement issues.

Expected outcome: The comparison distinguishes material/container problems from an analyzer-wide aspiration problem.

7. Check for Related Fluidics or Wash Conditions

Review normal analyzer status for concurrent fluid supply, waste, wash, or other conditions that could affect pipetting performance.

Expected outcome: Supporting fluidics are available and no broader fault is evident.

8. Perform Final Functional Verification

Run appropriate non-patient test material or required QC to confirm repeated aspiration occurs without errors and results meet laboratory acceptance requirements.

Expected outcome: Aspiration is stable and repeatable. Troubleshooting can stop when required verification passes.

If the Problem Persists

Sample condition, positioning, container compatibility, visible obstruction, and external fluidics conditions have been ruled out. The remaining issue may involve probe alignment, liquid-level sensing, pipetting mechanics, pressure sensing, tubing, valves, pumps, or other internal service-level components.

Remove the analyzer or affected testing from service, label it Out of Service, and arrange qualified repair or bench/on-site evaluation using Beckman Coulter documentation and approved test equipment. Do not bend probes, open fluidic assemblies, or make unauthorized mechanical adjustments.

Return the analyzer to service only after aspiration performance and required QC are successfully verified.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

A single successful aspiration does not prove reliability after repeated pipetting faults; verify repeatable operation before resuming patient testing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported repeated sample aspiration errors on the DxC 700 AU."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the affected specimen contained a visible clot that interfered with aspiration."

Resolution

What action was taken.

Example:
"The specimen was removed from the analyzer, laboratory staff prepared an acceptable replacement sample, and subsequent aspiration and QC operation were verified without error."

Helpful Details to Include (If Known)

Final Thought

Start with the material being aspirated before assuming a probe or pipettor failure. Verify volume, container, positioning, and fluidics first, stop when external troubleshooting limits are reached, and document successful repeatability before return to service.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide