Beckman Coulter GEM Premier 4000 / 5000 Series

Printer Feeds but the Report Is Blank or Missing Analytes

On this page

Asset Type

Blood Gas Analyzer

Manufacturer

Beckman Coulter

Model

GEM Premier 4000 / 5000 Series

What This Guide Helps With

Troubleshoots blank, incomplete, or missing printed results caused by paper, printer loading, report configuration, result suppression, or printer malfunction.

Step-by-Step Troubleshooting

1. Protect the Result-Reporting Workflow

Confirm that patient results remain available through the analyzer display, LIS, or another approved reporting method before troubleshooting the printer.

Do not delay urgent patient testing solely because the local printer is unavailable if an approved alternate reporting workflow exists.

If the printed report is the required documentation method and no alternate is available, remove the analyzer from that workflow until reporting reliability is restored.

Expected outcome: Patient results remain accessible through a verified reporting path while printer troubleshooting is performed.

2. Determine Whether the Problem Is Printing or Result Generation

Review the completed test on the analyzer display before printing.

Confirm whether the supposedly missing analytes:

This distinction prevents a measurement or configuration problem from being incorrectly treated as a printer fault.

Expected outcome: It is clear whether the failure affects the printer output, report configuration, or the actual test result.

If the analytes are absent from the analyzer result itself, troubleshoot the measurement or configuration issue rather than the printer.

3. Inspect the Printer Paper

Confirm that the correct approved paper is installed and loaded in the proper orientation.

A thermal printer can feed normally while producing a blank page if thermal paper is reversed.

Inspect for:

Expected outcome: Correct paper is installed in the proper orientation and feeds smoothly.

If reloading the correct paper produces a complete readable report, troubleshooting can stop after verification.

4. Inspect the Accessible Printer Path

Open only the normal operator-accessible printer area.

Check for:

Do not disassemble the printer mechanism.

Expected outcome: The accessible paper path is clear and the printer closes and feeds correctly.

If clearing an obstruction restores complete printing, verify several appropriate reports and stop troubleshooting.

5. Compare Blank Printing With Incomplete Printing

If the entire page is blank, focus on paper orientation/type and the printer's thermal printing function.

If text prints but certain analytes are absent, compare the displayed analyzer result with the report.

Determine whether the missing information is consistent across multiple reports.

Expected outcome: The problem is categorized as either failure to create an image on the paper or failure to include specific result fields.

This determines whether to continue with report configuration checks or escalate a printer hardware problem.

6. Check Report and Print Configuration

Using authorized normal configuration screens, verify the selected report format and applicable printing options.

Check for recent changes to:

Do not alter restricted configuration or service parameters without authorization.

Expected outcome: The selected report configuration is appropriate and includes the expected analytes.

If correcting an authorized report setting restores complete reports, troubleshooting can stop after verification.

7. Confirm Whether Missing Analytes Were Suppressed by the Analyzer

Review the result screen for flags, failed measurements, QC restrictions, unavailable tests, or other indications that an analyte was never released for reporting.

A printer cannot print a result that the analyzer did not generate or release.

Expected outcome: Every analyte expected on the printout is confirmed as an available, reportable analyzer result.

If the analyzer itself is suppressing the analyte, address that measurement or QC issue separately.

8. Print a Known Complete Report

Select an existing appropriate result or approved test record containing known reportable values and produce another printout.

Compare:

Expected outcome: The printer reproduces the full report consistently.

If the report is complete and readable, troubleshooting can stop.

9. Perform Final Functional Verification

Before closing the work order, verify that:

Expected outcome: Complete, readable reports are produced reliably.

If these conditions are met, troubleshooting is complete.

If the Problem Persists

If approved paper, loading orientation, accessible printer path, analyzer results, and report configuration have been verified but printing remains blank or incomplete, common external causes have been ruled out.

The remaining problem may involve the printer assembly, thermal print mechanism, internal connection, printer control electronics, software/report generation, or another service-level issue.

If reliable result reporting cannot be maintained through an approved alternate method, the analyzer should be:

Do not disassemble the printer mechanism or perform board-level repair without appropriate service authorization.

After repair, verify multiple complete reports and confirm required analyzer testing before return to service.

Knowing when a printer problem has progressed beyond paper and configuration checks is proper troubleshooting.

Clinical Use Tip

Always compare a report with the analyzer's displayed result before treating a missing printed analyte as a measurement failure.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory reported that the printer fed paper normally but all printed reports were blank."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the thermal paper was installed with the non-printing side facing the print mechanism."

Resolution

What action was taken.

Example:
"Paper was reloaded in the correct orientation, multiple complete reports were verified against displayed results, and the analyzer remained in service."

Helpful Details to Include (If Known)

Final Thought

Separate result-generation problems from printer problems before replacing hardware. Verify the displayed result, paper, loading, report configuration, and accessible printer path first, then escalate persistent failures and document a complete final reporting check.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide