On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoot AC supply, power connections, charging indications, battery state, transport setup, and runtime problems before suspecting an internal charging-system failure.
Step-by-Step Troubleshooting
1. Protect the Patient Before Testing Battery Operation
Do not perform transport or battery troubleshooting while the patient depends on an RI-2 with uncertain runtime.
Use stable AC power or another verified device as appropriate before removing external power.
Expected outcome: Patient therapy is protected from unexpected power loss.
2. Confirm the Reported Battery Problem
Determine whether the problem is:
- Battery does not appear to charge
- Battery runtime is unexpectedly short
- Unit shuts down when AC is disconnected
- Transport operation cannot be started
- Battery indication is inconsistent
- The device was recently stored or left disconnected
Record any exact display indications.
Expected outcome: The battery complaint is clearly defined and reproducible.
3. Verify the AC Power Source
Confirm the device is connected to a known-good appropriate outlet.
If permitted by facility practice, verify the outlet with another suitable device or electrical tester.
Expected outcome: Reliable AC power is available to the RI-2.
4. Inspect the Power Cord and Connection
Inspect the external power cord, plug, strain relief, and device-side connection for:
- Cuts
- Fraying
- Bent components
- Loose connection
- Heat damage
- Contamination
Remove damaged power components from service.
Expected outcome: The AC connection is secure and physically intact.
5. Verify Charging Indication
Connect the device to AC power and observe the available battery or charging indication.
Do not assume that an illuminated display confirms the battery is charging.
Expected outcome: The device recognizes AC power and shows an appropriate charging state.
6. Allow Appropriate Charging Opportunity
If the battery may be depleted from storage or transport use, allow it to charge in accordance with approved manufacturer guidance before evaluating runtime.
Do not invent a charging duration or expected runtime.
Expected outcome: The battery receives an appropriate charge before further testing.
7. Test Transition From AC to Battery
With the device removed from patient use and configured for bench testing, observe its behavior when external AC power is removed.
Verify that the device remains powered and indicates battery operation appropriately.
Expected outcome: The RI-2 transitions to battery operation without an unexpected shutdown.
8. Evaluate Battery Runtime
Using approved testing practices, evaluate whether battery operation is consistent enough for intended transport use.
If the unit immediately shuts down or runtime is clearly inadequate compared with applicable service criteria, do not return it for transport use.
Expected outcome: Battery performance meets applicable return-to-service requirements.
9. Inspect for Environmental or Storage Factors
Check whether the unit has been exposed to:
- Extended storage without charging
- Excessive heat or cold
- Fluid contamination
- Physical impact
- Damaged external connections
Document relevant conditions without assuming they are the cause.
Expected outcome: External conditions affecting battery performance are identified or ruled out.
10. Escalate Charging or Runtime Failure
If AC power is verified and the unit still will not charge, transition to battery, or provide acceptable runtime, remove it from service for qualified battery and charging-system evaluation.
Expected outcome: The RI-2 is not returned for transport use with unreliable backup power.
If the Problem Persists
Outlet problems, power-cord issues, charging opportunity, and obvious environmental factors have been ruled out.
The remaining cause may involve the battery, charging circuitry, power-management system, internal connection, or service-level calibration or configuration.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate manufacturer documentation and approved test equipment
- Repaired or configured only by qualified personnel
After repair or battery replacement, perform appropriate electrical, charging, runtime, and functional verification before return to service.
Knowing when to stop external troubleshooting and escalate is proper troubleshooting.
Clinical Use Tip
Verify adequate battery operation before transport begins; a rapid infuser that works only while plugged into AC is not ready for battery-dependent transport.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the RI-2 shut down when disconnected from AC power for transport."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering confirmed that AC power was normal but the battery could not sustain device operation."
Resolution
What action was taken.
Example:
"The unit was removed from service and sent for qualified battery-system evaluation and required return-to-service testing."
Helpful Details to Include (If Known)
- Battery indication
- Charging indication
- AC outlet verification
- Power-cord condition
- Whether unit operates normally on AC
- Behavior when AC is disconnected
- Storage history
- Runtime test result
- Any physical or fluid damage
- Final device status
Final Thought
Battery troubleshooting starts with verified AC power and safe bench testing. Never depend on an uncertain battery for patient transport, and escalate when reliable charging and runtime cannot be demonstrated.
That is successful troubleshooting.