Berchtold Chromophare Surgical Light

Arm Balance or Suspension Drift Issues

On this page

Asset Type

Surgical Light

Manufacturer

Berchtold

Model

Chromophare Series

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting situations where a Berchtold Chromophare surgical light arm drifts or does not stay in the intended position. Symptoms may include the light slowly dropping or rising on its own, difficulty in positioning, or uneven support when moving the arm. The focus is on external, verifiable causes before considering internal adjustments or repair.

Step-by-Step Troubleshooting

Verify Power and Control Settings

Why: Power fluctuations or inactive motorized assists can affect suspension tension.

Inspect Mechanical Locks and Brakes

Why: Loose brakes can cause the arm to drift even if internal balance is correct.

Check Arm Joints and Pivots

Why: Dirt or wear at pivot points can create uneven tension and drift.

Assess Counterbalance Mechanisms

Why: Improperly seated counterbalance components are a common cause of drift.

Test Movement Under Load

Why: Confirms whether the issue is balance-related or external obstruction.

Check for Environmental Factors

Why: External forces can mimic suspension failure.

If the Problem Persists

External adjustments have been ruled out, and drift continues:

Clinical Engineering principle: knowing when to stop is proper troubleshooting. Avoid internal spring or motor repairs unless certified.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"The surgical light arm drifts downward when positioned; it does not stay in place."

Cause

What was observed during troubleshooting.

Example:
Observed loose brake knobs and slight misalignment of pivot points causing uneven suspension tension.

Resolution

What action was taken.

Example:
Tightened all brake knobs and inspected pivot points. Noted that arm drift persists; light removed from service and labeled Out of Service for manufacturer evaluation.

Helpful Details to Include

Final Thought

Patient safety comes first: never attempt internal spring or motor repairs unless trained and certified. Logic-based troubleshooting and accurate documentation help maintain safe surgical environments and avoid unnecessary downtime. Escalating appropriately prevents further equipment damage and ensures consistent operating room functionality.

That is successful troubleshooting.