bioMerieux VITEK 2 Compact

Calibration Fails or Will Not Complete

On this page

Asset Type

Microbiology Identification System

Manufacturer

bioMerieux

Model

VITEK 2 Compact

What This Guide Helps With

Use this guide when a required calibration or verification process fails, stops unexpectedly, or cannot be completed successfully.

Step-by-Step Troubleshooting

1. Protect Test Validity

Do not use an analyzer for patient testing when a required calibration or associated verification has failed and the system's performance cannot be assured.

Laboratory staff should redirect testing through an alternate validated process.

Expected outcome: Patient results are not produced from an analyzer with unresolved calibration status.

2. Confirm What Procedure Failed

Determine exactly which calibration or verification process was attempted and whether the failure:

Record the exact failure message.

Expected outcome: The affected procedure and failure point are clearly documented.

3. Verify Analyzer Readiness

Confirm that the analyzer:

Calibration should not be used to mask an unrelated system fault.

Expected outcome: The analyzer is otherwise stable and ready for the procedure.

4. Verify Calibration Materials

Have qualified laboratory personnel confirm that the correct approved materials are being used and that they are acceptable for the procedure.

Check externally for:

Clinical Engineering should not determine laboratory acceptability beyond the scope of technical support.

Expected outcome: Appropriate material is available and acceptable for use.

5. Check Loading and Positioning

Verify that required materials, cards, racks, or carriers are correctly positioned and fully seated.

Inspect the loading area for contamination, obstruction, or damaged accessories.

Expected outcome: Calibration materials are positioned normally and detected by the system.

6. Verify External Environmental Conditions

Check for facility conditions that may affect system stability, including:

Do not invent or apply environmental limits not provided by the manufacturer.

Expected outcome: No obvious environmental problem is present.

7. Repeat the Procedure Only When Appropriate

If the failure appears attributable to a correctable external cause and laboratory personnel confirm the material is suitable, repeat the approved procedure once.

Do not repeatedly consume calibration materials or force repeated attempts when the same failure recurs.

Expected outcome: Calibration completes successfully after the external cause is corrected.

If calibration completes and subsequent required checks are acceptable, troubleshooting can stop.

8. Compare With Known-Good Materials When Authorized

If the laboratory has another verified set of appropriate materials, use that comparison to determine whether the failure follows the material or remains with the analyzer.

Expected outcome: The comparison separates material-related issues from analyzer-related issues.

9. Perform Final Functional Verification

After successful calibration, confirm that:

Expected outcome: Calibration and required post-calibration checks are complete and acceptable.

If the Problem Persists

If analyzer readiness, materials, loading, environment, external connections, and an appropriate repeat attempt have been verified, common external causes have been ruled out.

The remaining problem may involve a sensor, optical or mechanical condition, internal control issue, software problem, stored calibration data, or another service-level fault.

The device should be:

Do not make unauthorized internal adjustments to force calibration to pass. Knowing when to stop external troubleshooting is proper troubleshooting.

Complete required calibration, QC, and laboratory acceptance before return to service.

Clinical Use Tip

A failed calibration is a test-validity issue; do not return the analyzer to patient testing merely because it otherwise appears operational.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported that calibration on the bioMerieux VITEK 2 Compact repeatedly failed before completion."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found that the calibration material carrier was not fully seated because of debris in the loading area."

Resolution

What action was taken.

Example:
"Cleaned the accessible loading area using approved methods, verified proper carrier seating, and laboratory staff completed calibration and required verification successfully."

Helpful Details to Include (If Known)

Final Thought

Treat calibration failure as a reason to stop patient testing, verify external conditions and materials first, and escalate when successful calibration cannot be restored without service-level intervention.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide