On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoot intermittent freezes, unexpected restarts, subsystem dropouts, or communication loss caused by connections, power, environment, peripherals, or service-level faults.
Step-by-Step Troubleshooting
1. Protect the Patient and Procedure
If the system freezes, restarts, or loses communication during patient care, stop relying on it immediately. Coordinate a safe procedure pause or alternate clinical workflow as appropriate.
Expected outcome: Patient care no longer depends on an unstable angiography system.
2. Characterize the Event
Determine what froze or restarted, which components remained operational, whether communication returned automatically, and what was happening immediately before the event. Record all displayed messages and timestamps.
Expected outcome: The event is described precisely enough to distinguish a whole-system restart from an isolated subsystem or communication loss.
3. Check for Power Disturbance
Verify facility power status and inspect accessible power connections. Determine whether other room equipment experienced a simultaneous interruption.
Expected outcome: An obvious external power disturbance is identified or ruled out.
4. Inspect External Interconnects
Check accessible network, display, console, control, detector-interface, and other external system connections for looseness, damage, strain, or movement-related interruption.
Expected outcome: All external communication paths are securely connected.
5. Check Environmental Conditions
Look for blocked ventilation, excessive dust at accessible vents, nearby heat sources, liquid exposure, unusual room conditions, or evidence that equipment is overheating. Do not open covers to inspect internal cooling components.
Expected outcome: No correctable external environmental condition is contributing to instability.
6. Check External Peripherals and Accessories
Determine whether the issue began after connecting or changing an external accessory, peripheral, network device, or removable media. Disconnect only nonessential approved peripherals when safe and appropriate to isolate the condition.
Expected outcome: A problematic external peripheral is identified or ruled out.
7. Perform a Controlled Restart
After patient care is protected, use the normal approved shutdown and restart process. Do not repeatedly power-cycle a system displaying signs of overheating, electrical damage, smoke, or unusual odor.
Expected outcome: The system completes startup and all expected subsystems reconnect.
8. Observe System Stability
Allow the system to operate through representative non-patient functions and normal equipment positioning. Watch for renewed freezing, communication loss, or unexpected restart.
Expected outcome: The system remains stable without recurring interruption.
9. Verify Communication and Major Functions
Confirm operator displays, controls, system components, and required network or subsystem communication paths are restored. Perform applicable return-to-service testing.
Expected outcome: The system remains fully responsive and connected. If the event was caused by an identified external condition and does not recur, troubleshooting can stop.
10. Escalate Any Recurrent Instability
If the freeze, restart, or communication loss repeats or no external cause is confirmed, remove the system from service.
Expected outcome: An intermittently unstable angiography system is not returned to patient use without further evaluation.
If the Problem Persists
External power, connections, peripherals, environment, and controlled restart have been evaluated. Remaining causes may include internal power instability, cooling problems, software or operating-system faults, storage problems, subsystem communication hardware, network infrastructure, configuration, or another service-level condition.
Remove the system from service, label it Out of Service, and arrange qualified manufacturer or service evaluation using approved documentation and diagnostic equipment. Preserve available event information or logs according to facility procedure, especially if the event occurred during patient care. Do not perform unauthorized internal repairs or software changes.
Return to service only after the cause has been addressed and the system demonstrates stable operation during appropriate functional testing. Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
An intermittent system freeze is a failure even if operation returns on its own; do not continue a procedure based solely on temporary recovery.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the Canon Alphenix Sky+ intermittently lost communication with an operator display during system positioning."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found an external display communication cable loose enough to interrupt communication when the connected equipment moved."
Resolution
What action was taken.
Example:
"Clinical Engineering secured the connection, exercised the system through normal positioning, verified stable display communication and system operation, and returned the unit to service."
Helpful Details to Include (If Known)
- Exact component that froze or restarted
- Timestamp of the event
- Activity occurring immediately beforehand
- Error or communication message
- Whether other equipment lost power
- External cable and connection condition
- Peripheral devices attached
- Environmental observations
- Whether movement reproduced the problem
- Restart result
- Results of stability testing
- Final system status
Final Thought
Protect the patient immediately when system stability is uncertain, investigate power and external communication paths before assuming internal failure, preserve useful event details, and escalate recurrent instability appropriately.
That is successful troubleshooting.