Canon Aquilion ONE / PRISM Edition

Image Quality Is Poor or Artifacts Are Present

On this page

Asset Type

CT Scanner

Manufacturer

Canon

Model

Aquilion ONE / PRISM Edition

What This Guide Helps With

Addresses poor CT images or artifacts caused by patient positioning, external objects, motion, accessories, protocol selection, or calibration-related conditions.

Step-by-Step Troubleshooting

1. Protect the Patient From Unnecessary Repeat Imaging

Do not repeatedly scan a patient to troubleshoot unexplained artifacts. If diagnostic image quality is unreliable, stop clinical use and use another verified imaging system when necessary.

Expected outcome: Unnecessary radiation exposure is avoided.

2. Confirm the Artifact or Quality Complaint

Review what staff observed. Determine whether the issue is noise, streaking, banding, shading, distortion, motion artifact, incomplete anatomy, or another visible problem.

Record which exams, anatomy, protocols, and patients are affected.

Expected outcome: The image-quality problem is clearly characterized.

3. Check Patient Positioning and Motion Factors

Verify patient centering, immobilization, scan positioning, and whether voluntary or involuntary motion could explain the image.

Expected outcome: Patient positioning and movement are either identified as the cause or ruled out.

If proper positioning eliminates the problem on subsequent clinically indicated imaging, troubleshooting may stop after verification.

4. Remove External Objects From the Scan Field

Inspect for cables, ECG leads, jewelry, positioning devices, supports, clothing hardware, blankets, injector components, or other dense objects in or near the scan field.

Expected outcome: No unnecessary external object contributes to artifact.

5. Check Accessories and Table Components

Inspect head holders, cushions, positioning aids, table pads, and removable accessories for damage, contamination, foreign material, or incorrect placement.

Expected outcome: External scanner accessories are correctly installed and artifact-free.

6. Verify Exam and Reconstruction Selection

Confirm that staff selected the intended protocol, scan range, reconstruction, field of view, and other normal clinical settings.

Do not alter restricted service or calibration values.

Expected outcome: Image appearance is not explained by an unintended exam or reconstruction selection.

7. Determine Whether the Problem Is Patient-Specific or System-Wide

Compare with prior acceptable studies or perform an approved phantom/QC check if within Clinical Engineering responsibility.

Do not expose additional patients for comparison.

Expected outcome: The issue is identified as patient/exam-specific or reproduced under controlled conditions.

8. Check Recent Environmental or Service Events

Ask whether the problem began after maintenance, power interruption, system restart, room temperature problem, hardware service, or software/configuration change.

Expected outcome: Any relevant event is identified for escalation or further testing.

9. Perform Approved Image-Quality Verification

Use appropriate approved phantom or quality-control methods to verify scanner performance.

Do not substitute subjective visual judgment for required QC testing.

Expected outcome: Image performance is either verified as acceptable or the artifact is reproducibly confirmed.

If image quality meets required criteria and the original external cause is corrected, troubleshooting is complete.

10. Escalate Persistent Image-Quality Problems

If artifacts remain during controlled testing, remove the scanner from service or restrict use according to site policy. Do not attempt detector calibration, X-ray tube adjustment, or internal component repair outside approved service procedures.

Expected outcome: Unverified image-quality performance is not used for patient diagnosis.

If the Problem Persists

Common external causes have been ruled out. Remaining possibilities may include detector calibration, acquisition electronics, X-ray generation, reconstruction processing, mechanical alignment, environmental influences, or service-level configuration.

The CT scanner should be:

Required image-quality and radiation-performance testing must be successfully completed before unrestricted clinical use resumes.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Do not use patient examinations as repeated test scans when artifact cause is unknown; use appropriate phantom or QC methods instead.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"CT staff reported streak artifacts appearing on head studies."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found a removable positioning accessory containing a dense external object in the scan field."

Resolution

What action was taken.

Example:
"Removed the interfering object, inspected the accessory, completed approved image-quality verification, and confirmed the artifact was no longer present."

Helpful Details to Include (If Known)

Final Thought

Avoid unnecessary repeat exposure, rule out patient positioning and external objects first, verify image quality objectively, and escalate persistent artifacts for qualified evaluation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide