Canon Aquilion Series

Tube or Generator Errors

On this page

Asset Type

CT Scanner

Manufacturer

Canon

Model

Aquilion Series

What This Guide Helps With

Tube, generator, exposure, or high-voltage errors that prevent scan initiation or interrupt CT operation.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Stop the scan attempt and confirm the patient is safe before troubleshooting.

Action:

Expected Outcome:
Patient exposure risk is controlled and the CT scanner is not repeatedly forced into failed exposure attempts.

2. Verify the Reported Error

Confirm the exact error message, code, or system behavior.

Action:

Expected Outcome:
The reported tube or generator issue is confirmed and documented accurately.

3. Check for Active Patient or Exam Status Issues

Make sure the system is not locked in a failed or incomplete exam state.

Action:

Expected Outcome:
If the issue was caused by an incomplete exam state or software lockout, the system may return to normal operation.

If this resolves the issue, document the finding and stop.

4. Check Emergency Stop and Room Safety Interlocks

Exposure may be inhibited if a safety circuit is active.

Action:

Expected Outcome:
All safety interlocks are normal and exposure is not being blocked externally.

If this resolves the issue, document the reset or interlock correction and stop.

5. Confirm Stable Facility Power

Tube and generator faults can occur if incoming power is unstable or interrupted.

Action:

Expected Outcome:
No obvious external power issue is present.

If a facility power issue is found, involve Facilities or the appropriate electrical support team before further scanner troubleshooting.

6. Perform a Controlled System Restart if Appropriate

Temporary software or communication faults may clear after a proper restart.

Action:

Expected Outcome:
The scanner boots normally and tube/generator errors clear if caused by a temporary communication or software state issue.

If this resolves the issue, document the restart and successful return to service.

7. Check Tube Warm-Up Status

Tube or generator errors may occur if warm-up is incomplete, skipped, or interrupted.

Action:

Expected Outcome:
Warm-up completes without fault and the system returns to ready status.

If warm-up fails again, stop repeated attempts and escalate.

8. Check for Tube Heat or Cooling-Related Warnings

The system may inhibit exposure if tube heat limits or cooling conditions are outside safe range.

Action:

Expected Outcome:
If the fault was heat-related, the system may recover after cooling and returning to normal operating temperature.

If the same error returns after cooldown, remove the scanner from service and escalate.

9. Inspect External Gantry and Cabinet Conditions

Look for obvious signs of overheating, electrical odor, or abnormal operation.

Action:

Expected Outcome:
No obvious external hazard is present.

If unusual heat, smell, smoke, or electrical noise is present, stop immediately, remove the scanner from service, and escalate urgently.

10. Check Operator Console Communication

Some generator or tube messages may be triggered by communication loss between the console, gantry, and generator subsystem.

Action:

Expected Outcome:
The console and scanner subsystems communicate normally.

If communication errors continue with tube/generator faults, escalate for service evaluation.

11. Attempt Only One Controlled Functional Retest

After external checks and a proper restart, perform a limited retest only if the system appears safe.

Action:

Expected Outcome:
System completes readiness testing without exposure faults.

If the error returns, stop troubleshooting and remove the scanner from service.

If the Problem Persists

If tube or generator errors continue after checking safety interlocks, power, system state, warm-up status, cooling conditions, and controlled restart, common external causes have been ruled out.

The issue is likely related to an internal tube, high-voltage generator, inverter, cable, cooling, control, or communication fault.

The device should be removed from service, labeled Out of Service, and escalated for Canon service, vendor repair, or qualified CT bench/system evaluation.

Knowing when to stop is proper troubleshooting. CT tube and generator systems involve high voltage, radiation exposure control, and critical safety circuits. Do not continue repeated exposure attempts or perform internal repair without proper training and authorization.

Clinical Use Tip

Do not troubleshoot tube or generator errors with a patient on the table unless the troubleshooting is limited to safe observation and immediate removal from service. Move the patient to another scanner or reschedule the exam before performing restart, warm-up, or retest steps so therapy continuity and patient care are maintained.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"CT staff reported that the Canon Aquilion scanner displayed a tube/generator error during scan initiation and would not complete exposure."

Cause

What was observed during troubleshooting.

Example:
"Verified recurring exposure inhibit after checking emergency stops, room interlocks, system status, cooling condition, and facility power with no external cause identified."

Resolution

What action was taken.

Example:
"Removed CT scanner from service, labeled unit Out of Service, documented error behavior, and escalated to Canon service for tube/generator system evaluation."

Helpful Details to Include (If Known)

Final Thought

Tube and generator errors on a Canon Aquilion CT scanner should be handled carefully because they involve radiation control, high voltage, and patient safety. Clinical Engineering should rule out external causes first, avoid repeated failed exposure attempts, and escalate once the issue points to the internal tube or generator system. Clear CCR documentation helps the vendor or imaging specialist continue troubleshooting efficiently.

That is successful troubleshooting.