On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Error: 11 Flash Read Error indicates a failure in reading internal memory, which can prevent the pump from starting or functioning correctly. This can result in complete loss of feeding capability and requires prompt evaluation to ensure patient safety.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Immediately remove the pump from patient use
- Transition feeding to a backup pump or method
- Confirm feeding continues without interruption
Verify the Reported Error and Behavior
- Power on the pump and observe the display
- Confirm Error: 11 Flash Read Error is shown
- Note whether it appears at startup or during operation
- Check if the device is frozen or non-functional
Perform a Power Cycle
- Turn the pump OFF
- Disconnect from AC power
- Remove and reinsert the battery if accessible
- Wait 30–60 seconds
- Reconnect power and restart the pump
Confirm Power Source Stability
- Plug the device into a known-good outlet
- Verify the power adapter connection is secure
- Observe for consistent startup behavior
Inspect for Physical or Environmental Issues
- Check for signs of liquid intrusion
- Inspect housing for cracks or impact damage
- Ensure the device is within normal temperature conditions
If the Problem Persists
If Error: 11 Flash Read Error continues after completing all external troubleshooting steps, internal causes have been confirmed.
- Remove the device from service
- Label the device as Out of Service
- Send the unit for repair or control board replacement
Clinical Use Tip
Never attempt to continue using a pump that intermittently clears this error. Memory-related faults can lead to unpredictable operation and unsafe feeding delivery. Always transition the patient to a backup device first.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Pump displaying Error: 11 Flash Read Error and will not start."
Cause
What was observed during troubleshooting.
Example:
"Internal flash memory read failure confirmed after repeated power cycles; error persisted at startup."
Resolution
What action was taken.
Example:
"Removed device from service and sent for repair due to suspected control board failure."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
Error: 11 Flash Read Error is a strong indicator of internal memory failure. Efficient troubleshooting focuses on verifying the error, ruling out external causes, and escalating quickly to protect patient safety and reduce downtime.
That is successful troubleshooting.