Cardinal Health Kangaroo ePump

Error: 14 Timeout during downstream occlusion check (rotor signal failure)

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo ePump

What This Guide Helps With

Error 14 indicates a failed downstream occlusion check due to missing rotor feedback or motor-related sensing failure. This can cause the pump to stop operation and prevents safe enteral feeding delivery.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Power Cycle the Pump

Inspect Feeding Set Installation

Check for Mechanical Obstruction

Swap Feeding Set

Observe Motor Operation

If the Problem Persists

If Error 14 continues after completing all external troubleshooting steps:

Clinical Use Tip

Never troubleshoot or rely on a feeding pump displaying active errors while connected to a patient. Always transition to a backup device first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pump displaying Error 14 and unable to initiate feeding; no motor movement observed."

Cause

What was observed during troubleshooting.

Example:
"Rotor signal failure during downstream occlusion check; no motor activity detected despite proper setup."

Resolution

What action was taken.

Example:
"Power cycled unit, reloaded feeding set, inspected rotor area, and replaced tubing with no resolution; removed from service and sent for repair."

Helpful Details to Include (If Known)

Final Thought

Error 14 points to a failure in rotor detection or motor feedback. Always rule out setup and mechanical issues first, then escalate appropriately. Patient safety and timely removal from service are critical.

That is successful troubleshooting.