Cardinal Health Kangaroo OMNI

Alarm: Cassette Dislodged

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo OMNI

What This Guide Helps With

This guide helps troubleshoot the Cassette Dislodged alarm on the Cardinal Health Kangaroo OMNI feeding pump. This alarm may occur when the cassette is not fully seated, the feeding set cassette is damaged or misaligned, contamination prevents proper loading, or the pump cannot reliably detect the cassette.

A Cassette Dislodged alarm can interrupt feeding or flushing and should be treated as a delivery interruption until the setup is verified safe.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Remove the Pump From Active Patient Use Before Reloading

Reload the Cassette

Check for Simple Loading Obstructions

Inspect the Feeding Set Cassette

Replace the Kangaroo OMNI Feeding Set

Check for Pump-Specific Behavior

Stop When Resolved

If the Problem Persists

If the Cassette Dislodged alarm continues after proper cassette loading, inspection, cleaning of obvious external contamination, and replacement of the Kangaroo OMNI feeding set, external and common causes have been ruled out.

A persistent Cassette Dislodged alarm may indicate a pump-side cassette detection, loading mechanism, or internal sensing issue.

Remove the pump from service, power it off, label it Out of Service, and send it for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device first.

For cassette-loading alarms, do not repeatedly reload or prime the cassette on an active patient setup without clinical awareness. If feeding or flushing is interrupted, transfer therapy to another verified pump rather than continuing to clear the alarm.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Kangaroo OMNI feeding pump reported with Cassette Dislodged alarm. Staff stated cassette alarm occurred after loading feeding set and pump would not continue feeding."

Cause

What was observed during troubleshooting.

Example:
"Cassette was not fully seated in the pump. Alarm cleared after cassette was removed, reloaded, and accepted by the pump."

Resolution

What action was taken.

Example:
"Reloaded cassette, pressed Continue, verified Cassette Dislodged alarm cleared, and confirmed pump operated normally during setup. Device returned to service."

Helpful Details to Include (If Known)

Final Thought

Cassette Dislodged alarms should be approached as real delivery interruptions until proven otherwise. Start with patient safety, confirm the alarm, reload the cassette correctly, inspect the feeding set, and replace the set before suspecting pump failure. If the alarm remains after basic external checks, remove the pump from service and escalate appropriately. Clear CCR documentation supports safe return to service or repair follow-up.

That is successful troubleshooting.