Cardinal Health Kangaroo OMNI

Alarm: Cassette Error

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo OMNI

What This Guide Helps With

This guide helps troubleshoot the Cassette Error alarm on the Cardinal Health Kangaroo OMNI feeding pump. This alarm may occur when the pump detects an unusual cassette operating condition, a cassette loading problem, disposable feeding set irregularity, contamination near the cassette area, or a pump-side cassette detection issue.

A Cassette Error may interrupt feeding delivery or prevent the pump from running reliably. Because enteral feeding accuracy and continuity matter, troubleshooting should begin with patient safety and simple external checks before removing the pump from service.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Remove the Cassette and Inspect the Feeding Set

Correct Any Obvious Cassette or Tubing Irregularities

Reload the Cassette Properly

Check for External Contamination or Moisture

Replace the Feeding Set

Test Pump Operation With the New Set

Isolate Device-Related Failure

If the Problem Persists

If the Cassette Error continues after correcting cassette irregularities, cleaning visible external contamination, reloading the cassette, and replacing the feeding set, external and common disposable-set causes have been ruled out.

Remove the pump from service, label it Out of Service, and send it for repair or bench evaluation.

Persistent cassette alarms may indicate an internal cassette detection, sensor, door/loading, or operating mechanism issue. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device first.

For a Cassette Error alarm, do not repeatedly reload or restart the same patient setup if feeding delivery is interrupted or uncertain. Use a known-good pump and evaluate the suspect pump separately.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Kangaroo OMNI feeding pump reported with Cassette Error alarm during feeding. Staff stated pump alarmed and feeding delivery was interrupted."

Cause

What was observed during troubleshooting.

Example:
"Cassette removed and inspected. Tubing/cassette irregularity suspected. Alarm cleared after cassette was reloaded and feeding set was replaced."

Resolution

What action was taken.

Example:
"Replaced feeding set, reloaded cassette, primed, and verified pump operation without recurring Cassette Error. Device returned to service."

Helpful Details to Include (If Known)

Final Thought

Cassette Error troubleshooting should begin with patient safety, then move through visible cassette condition, proper loading, contamination checks, and feeding set replacement. If the alarm continues after the disposable set is ruled out, the pump should be removed from service for repair evaluation. Clear documentation helps show whether the problem was corrected externally or requires deeper service.

That is successful troubleshooting.