Cardinal Health Kangaroo OMNI

Alarm: Dead Battery Error

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo OMNI

What This Guide Helps With

This guide helps troubleshoot a Dead Battery Error on the Cardinal Health Kangaroo OMNI feeding pump when battery failure is imminent and feeding or flushing may stop due to battery loss. This alarm should be treated as an urgent power condition because the pump may stop delivering fluid if the battery can no longer support operation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Connect the Pump to Wall Power Immediately

Restart Pump Operation

Check for Charging Response

Inspect External Power Components

Allow Adequate Charging Time

Test Battery Operation After Charging

Identify Battery Pack Failure

Stop When Resolved

If the Problem Persists

If the Dead Battery Error continues after confirming the wall outlet, power adapter, power connection, and charging response, external and common causes have been ruled out.

A persistent dead battery condition strongly indicates battery pack failure or an internal charging or power issue.

The device should be removed from service, labeled Out of Service, and sent for qualified battery replacement or bench evaluation.

Do not continue cycling power or repeatedly restarting the pump for patient use.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a dead battery alarm on an active patient while relying on the pump for feeding or flushing. Connect to AC power immediately or move the patient to a backup device first. A pump with unresolved battery failure should not be used for transport or unplugged operation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Kangaroo OMNI feeding pump reported Dead Battery Error during use. Staff stated pump was near battery loss and feeding or flushing operation had to be restarted."

Cause

What was observed during troubleshooting.

Example:
"Battery was fully depleted or battery pack failed to recover charge after confirmed AC power connection. External power source and adapter connection were checked."

Resolution

What action was taken.

Example:
"Connected pump to known-good wall outlet, verified power adapter seating, restarted operation, and monitored charging response. Pump continued to show dead battery condition and was removed from service for battery replacement or bench evaluation."

Helpful Details to Include (If Known)

Final Thought

A Dead Battery Error on the Kangaroo OMNI should be treated as an immediate patient-care risk because delivery may stop when battery loss occurs. Clinical Engineering should verify the alarm, confirm AC power and adapter integrity, restart operation safely, and determine whether the battery can recover. If the battery cannot charge or support operation, remove the pump from service and escalate for qualified repair. Clear CCR documentation protects the patient, the device history, and the next technician.

That is successful troubleshooting.