Cardinal Health Kangaroo OMNI

Alarm: Flush Error

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo OMNI

What This Guide Helps With

This guide helps troubleshoot Flush Error alarms on the Cardinal Health Kangaroo OMNI feeding pump. This alarm may occur when the pump detects a flush delivery problem caused by an empty flush bag, kinked tubing, clog, cassette issue, improper priming, or a flush occlusion condition.

A Flush Error can interrupt ordered flush delivery and may affect hydration, tube patency, medication delivery plans, or feeding workflow if not corrected promptly.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Understand What the Alarm Means

Check the Flush Bag

Inspect the Flush Tubing Path

Verify Flush Tubing Was Properly Primed

Check Cassette Seating and Loading

Clean and Dry the Occlusion Sensor Region

Replace the Feeding Set if Needed

Test With a Known-Good Setup

If the Problem Persists

If the Flush Error continues after confirming the flush bag is filled, tubing is not kinked, the line is properly primed, the cassette is correctly loaded, the occlusion sensor area is clean and dry, and the feeding set has been replaced, external and common causes have been ruled out.

At that point, the persistent alarm may indicate a pump-side sensing or occlusion detection problem.

Remove the pump from service, label it Out of Service, and send it for bench evaluation or repair.

Use a different pump for patient care.

Clinical Use Tip

Never troubleshoot a Flush Error alarm on an active patient. Move the patient to a backup device first if continued feeding or flushing is needed.

Do not prime flush or feed tubing while connected to the patient. Disconnect the patient before priming and avoid relying on a feeding pump that has an unresolved Flush Error alarm.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Kangaroo OMNI feeding pump reported with Flush Error alarm during programmed flush. Staff stated flush would not complete."

Cause

What was observed during troubleshooting.

Example:
"Flush bag and tubing path inspected. Flush tubing was found kinked between the flush bag and pump, preventing proper flush delivery."

Resolution

What action was taken.

Example:
"Corrected tubing position, verified flush bag volume, primed flush tubing with patient disconnected, and confirmed flush completed without alarm. Device returned to service."

Helpful Details to Include (If Known)

Final Thought

A Flush Error should be approached as a real delivery-path problem until proven otherwise. Start with patient safety, verify the alarm, check the flush bag and tubing, confirm proper priming, inspect the cassette and sensor area, and replace the disposable set before suspecting the pump.

If the alarm persists after external causes are ruled out, remove the pump from service and document the findings clearly.

That is successful troubleshooting.