Cardinal Health Kangaroo OMNI

Alarm: Low Battery Caution

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo OMNI

What This Guide Helps With

This guide helps troubleshoot Low Battery Caution alarms on the Cardinal Health Kangaroo OMNI feeding pump when battery life is critically low but feeding or flushing may continue. This alarm should be treated as a time-sensitive power warning because therapy could stop if the battery fully discharges.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Connect the Pump to Wall Power

Confirm the Outlet Is Working

Inspect the Power Adapter Connection

Check the Power Adapter

Charge the Pump Fully

Verify Battery Operation After Charging

Stop When Resolved

If the Problem Persists

If the Low Battery Caution alarm remains while connected to a known-good outlet and known-good power adapter, or if the alarm returns shortly after a full 12-hour charge, external and common charging causes have been ruled out.

Persistent Low Battery Caution may indicate a failed battery, charging circuit issue, power adapter port problem, or internal power management failure.

Remove the pump from service, label it Out of Service, and send it for repair or bench evaluation.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a Low Battery Caution alarm on an active patient while the pump is relying on battery power. Connect to AC power immediately or transfer therapy to another pump before the battery fully depletes.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Cardinal Health Kangaroo OMNI feeding pump displayed Low Battery Caution alarm during feeding. Staff reported pump was still delivering fluid but battery warning indicated low remaining runtime."

Cause

What was observed during troubleshooting.

Example:
"Pump battery was critically low. Power adapter connection and wall outlet were checked. Adapter plug was confirmed fully seated in the pump power adapter port. No external power damage noted."

Resolution

What action was taken.

Example:
"Pump was connected to a known-good wall outlet and charged for at least 12 hours. Alarm cleared after charging. Pump passed basic power and operational verification and was returned to service."

Helpful Details to Include (If Known)

Final Thought

The Low Battery Caution alarm is a time-sensitive power warning, not an immediate pump-stop condition. Clinical Engineering should confirm the alarm, protect patient therapy, check the outlet and adapter connection, fully charge the pump, and escalate if the battery or charging system cannot be trusted. Clear CCR documentation supports safe return to service or appropriate repair escalation.

That is successful troubleshooting.