Cardinal Health Kangaroo OMNI

Alarm: Low Battery Notice

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo OMNI

What This Guide Helps With

This guide helps troubleshoot Low Battery Notice alarms on the Cardinal Health Kangaroo OMNI feeding pump when the pump has about 30 minutes of battery life remaining. The pump may continue feeding or flushing during this alarm, but the condition should be corrected before the battery becomes depleted and therapy is interrupted.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Connect the Pump to Wall Power

Verify the Wall Outlet

Inspect the Power Adapter and Cord

Allow Adequate Charging Time

Recheck Operation After Charging

Isolate Adapter Versus Pump or Battery Issue

Stop When Resolved

If the Problem Persists

If the Low Battery Notice continues while the pump is connected to a known-good outlet with a known-good adapter, external causes have been ruled out.

Persistent low battery alarms may indicate a failed or weak internal battery, faulty charging circuit, damaged power adapter port, or internal power management issue.

Remove the pump from service, label it Out of Service, and send it for repair or bench evaluation.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device first if therapy may be interrupted, if the pump cannot remain connected to AC power, or if battery reliability is uncertain.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Cardinal Health Kangaroo OMNI feeding pump displayed Low Battery Notice during feeding or flushing. Staff reported pump was low on battery with approximately 30 minutes of battery life remaining."

Cause

What was observed during troubleshooting.

Example:
"Pump battery was depleted. Power adapter connection and wall outlet were checked. Pump accepted wall power after adapter was fully seated and was placed on charge."

Resolution

What action was taken.

Example:
"Connected pump to known-good wall outlet, verified power adapter was fully inserted into pump power adapter port, and charged pump. Alarm cleared and pump operation verified. Returned to service."

Helpful Details to Include (If Known)

Final Thought

The Low Battery Notice is usually a straightforward power condition, but it still affects therapy reliability. Start with patient safety, confirm the alarm, check the outlet and adapter connection, allow a full charge, and escalate if the pump cannot maintain or accept battery charge.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide