Cardinal Health Kangaroo OMNI

Alarm: Patient Tube Blocked

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo OMNI

What This Guide Helps With

This guide helps troubleshoot the Patient Tube Blocked alarm on the Cardinal Health Kangaroo OMNI feeding pump. This alarm indicates that feeding solution is not being delivered as expected because of a downstream restriction between the pump and the patient.

Common causes include kinked tubing, closed clamps, tubing pinched by the bed or patient, patient-side obstruction, formula buildup, cassette loading issues, or contamination near the occlusion sensor region.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Expected outcome: The patient is not left connected to a pump that is not delivering feeding solution.

Why it matters: This alarm means the pump believes feeding solution is no longer being delivered to the patient due to a blockage between the pump and patient.

2. Verify the Reported Alarm and Behavior

Expected outcome: The Patient Tube Blocked alarm is confirmed and the timing is understood.

Why it matters: Understanding when the alarm appears helps separate a true downstream restriction from a setup, cassette, or sensing issue.

3. Inspect the Tubing Between the Pump and Patient

Expected outcome: Tubing from the pump to the patient is open and unrestricted.

Why it matters: This alarm specifically points to a downstream blockage between the pump and patient.

4. Confirm the Feeding Set Is Properly Positioned

Expected outcome: The pump can sense and move fluid through the set normally.

Why it matters: A poorly seated cassette or misrouted tube can create false pressure readings or actual flow restriction.

5. Clear the Alarm and Restart Feeding

Expected outcome: Feeding resumes without repeat downstream blockage alarms.

Why it matters: A cleared alarm after correcting the tubing path confirms an external downstream restriction was likely the cause.

6. Remove, Inspect, and Reload the Cassette

Expected outcome: The cassette is seated correctly and the occlusion sensor area is clean and dry.

Why it matters: Contamination or moisture near the occlusion sensing area can interfere with proper pressure detection.

7. Restart Feeding After Cassette Reload

Expected outcome: Feeding runs without repeated downstream blockage alarms.

Why it matters: This confirms whether the alarm was related to cassette seating or occlusion sensor area contamination.

8. Replace the Feeding Set

Expected outcome: The alarm clears if the original feeding set was restricted, damaged, or causing sensing problems.

Why it matters: The feeding set may have an internal restriction, kink, cassette defect, or formula buildup that is not visible externally.

9. Try a Different Pump if Needed

Expected outcome: The issue is isolated as pump-related or feeding-set/patient-line related.

Why it matters: Swapping the pump or set helps avoid replacing the wrong item or returning a faulty pump to service.

If the Problem Persists

If the Patient Tube Blocked alarm continues after checking the downstream tubing, clearing visible restrictions, cleaning and drying the occlusion sensor region, reloading the cassette, and replacing the feeding set, external and common causes have been ruled out.

Remove the pump from service.

Label it Out of Service and send it for repair or bench evaluation.

Persistent Patient Tube Blocked alarms may indicate an internal occlusion sensing, cassette interface, or pump mechanism problem.

Clinical Use Tip

Never troubleshoot a Patient Tube Blocked alarm on an active patient without clinical awareness. Do not repeatedly clear the alarm while the patient remains connected unless the tubing path has been verified open. Move the patient to a backup device first if feeding delivery cannot be restored quickly and safely.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Kangaroo OMNI feeding pump reported Patient Tube Blocked alarm during feeding. Staff stated feeding stopped and formula was not being delivered to the patient."

Cause

What was observed during troubleshooting.

Example:
"Downstream tubing between pump and patient was kinked, causing restricted flow and Patient Tube Blocked alarm."

Resolution

What action was taken.

Example:
"Cleared tubing restriction, verified cassette was seated correctly, pressed Continue, restarted feeding, and confirmed pump operated without recurring alarm. Returned pump to service."

Helpful Details to Include (If Known)

Final Thought

The Patient Tube Blocked alarm should be treated as a real delivery interruption until proven otherwise. Start with patient safety, inspect the tubing path from pump to patient, verify cassette seating, and clean the sensing area before suspecting internal failure. Good documentation confirms what was checked, what corrected the issue, and whether the pump was safe to return to service.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide