What This Guide Helps With

This guide helps troubleshoot the Pump Inactive alarm on the Cardinal Health Kangaroo OMNI feeding pump. This alarm appears when the pump has been without user input for more than 10 minutes.

The condition is usually a workflow timeout rather than a hardware failure, but it still requires confirmation that feeding or flushing has not been unintentionally delayed.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Confirm Whether the Pump Was Left Without Input

Press Continue

Verify the Pump Is Actually Running if Feeding Is Needed

Check for Repeated Inactivity Complaints

Inspect the Touchscreen and Controls if the Alarm Will Not Clear

Power Cycle Only if Clinically Safe

Stop When Resolved

If the Problem Persists

If the Pump Inactive alarm cannot be cleared, the screen is frozen, controls do not respond, or the pump repeatedly enters the alarm despite normal interaction, remove the pump from service.

At that point, external and common causes have been ruled out, including user inactivity, incomplete setup, pause status, and normal timeout behavior.

Label the device Out of Service and send it for repair or bench evaluation. The issue may involve touchscreen failure, control input failure, software lockup, or another internal device fault.

Clinical Use Tip

Never troubleshoot the Kangaroo OMNI on an active patient if feeding or flushing status is uncertain. Move the patient to a backup feeding pump first, then troubleshoot the reported alarm.

Do not assume the pump is actively feeding after clearing a Pump Inactive alarm. Always confirm the pump is running and delivering as intended before leaving the bedside.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Kangaroo OMNI reported with Pump Inactive alarm. Staff stated pump displayed inactivity screen after being left without input during setup."

Cause

What was observed during troubleshooting.

Example:
"Pump Inactive alarm verified. Alarm occurred because the pump had no user input for more than 10 minutes. Continue button responded normally and no hardware failure was found."

Resolution

What action was taken.

Example:
"Pressed Continue and pump returned to previous screen. Verified display and controls responded normally. Pump programmed and function checked successfully. Returned to service."

Helpful Details to Include (If Known)

Final Thought

The Pump Inactive alarm is usually a timeout reminder, not a pump failure. Clinical Engineering should still verify patient safety, confirm the pump status, check for unresponsive controls, and document clearly when the alarm is resolved by normal operation.

That is successful troubleshooting.