Cardinal Health Kangaroo OMNI

Alarm: Sensor Error 1

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo OMNI

What This Guide Helps With

This guide helps troubleshoot Sensor Error 1 on the Cardinal Health Kangaroo OMNI feeding pump. This alarm may be caused by occlusion sensor contamination, cassette tubing debris, feeding set issues, moisture, residue, or pump sensor degradation. Because Sensor Error 1 involves the occlusion detection system, the pump should not be relied on for feeding or flushing until the condition is resolved or the device is replaced.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Remove the Cassette

Inspect the Occlusion Sensor Region

Clean and Dry the Area

Reload the Cassette Correctly

Replace the Feeding Set

Isolate Pump vs. Disposable Set

If the Problem Persists

If Sensor Error 1 continues after the cassette area is cleaned and dried, the tubing inside the cassette has been checked, and the feeding set has been replaced, remove the pump from service.

Label the device Out of Service and send it for repair or bench evaluation.

At that point, external causes such as cassette loading, tubing debris, contamination, moisture, and feeding set failure have been ruled out. Persistent Sensor Error 1 may indicate an internal occlusion detection sensor issue or sensor degradation condition.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device first.

Do not keep restarting feeding on a pump with an unresolved Sensor Error 1 alarm. Transfer feeding to another Kangaroo OMNI pump before continuing patient therapy.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Kangaroo OMNI feeding pump reported Sensor Error 1 during feeding. Staff reported feeding stopped and alarm returned after restart."

Cause

What was observed during troubleshooting.

Example:
"Occlusion sensor region and cassette tubing inspected. Debris or contamination was found in the cassette sensor area, causing abnormal occlusion detection sensor behavior."

Resolution

What action was taken.

Example:
"Removed cassette, cleaned and dried occlusion sensor region using non-sharp cleaning method, inspected tubing inside cassette, reloaded cassette, pressed Continue, and verified feeding resumed without Sensor Error 1. Pump returned to service."

Helpful Details to Include (If Known)

Final Thought

Sensor Error 1 should be treated as a real sensor-related condition until proven otherwise. Start with patient safety, then check the cassette, tubing, contamination, moisture, and feeding set before suspecting pump failure. If the alarm continues after external causes are ruled out, removing the pump from service is the correct Clinical Engineering decision. Clear CCR documentation supports safe follow-up and repair tracking.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide