Cardinal Health Kangaroo OMNI

Alarm: Sensor Error 3

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo OMNI

What This Guide Helps With

This guide helps troubleshoot Sensor Error 3 on the Cardinal Health Kangaroo OMNI feeding pump when the pump detects possible force sensor degradation, cassette contact problems, residue, moisture, or feeding set issues that may prevent normal feeding operation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Remove the Feeding Set Cassette

Clean and Dry the Occlusion / Force Sensor Region

Reload the Cassette Correctly

Replace the Feeding Set

Check for Physical Damage or Contamination

Power Cycle and Retest

Isolate the Cause

If the Problem Persists

If Sensor Error 3 continues after cleaning the sensor region, reloading the cassette, replacing the feeding set, and power cycling the pump, external causes have been ruled out.

Remove the Kangaroo OMNI from service, label it Out of Service, and send it for repair or bench evaluation.

Persistent Sensor Error 3 may indicate internal force sensor degradation or sensor detection failure. Do not continue troubleshooting on the floor or return the pump to patient use with an unresolved sensor alarm.

Clinical Use Tip

Never troubleshoot Sensor Error 3 on an active patient. Move the patient to a backup feeding pump first, then continue troubleshooting the affected device only after patient feeding needs are safely supported.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Cardinal Health Kangaroo OMNI feeding pump reported Sensor Error 3 during power-up or cassette handling and prevented normal feeding operation."

Cause

What was observed during troubleshooting.

Example:
"Occlusion/force sensor region inspected and cleaned. Feeding set cassette was reloaded and replaced with a known-good set. Alarm persisted, indicating likely internal force sensor degradation or pump sensor hardware failure."

Resolution

What action was taken.

Example:
"Pump removed from patient use, labeled Out of Service, and sent for bench evaluation/repair. Clinical area advised to use a different Kangaroo OMNI pump."

Helpful Details to Include (If Known)

Final Thought

Sensor Error 3 should be treated as a meaningful sensor-related alarm, not just a nuisance message. Start with patient safety, then check cassette seating, residue, moisture, feeding set condition, and power behavior before suspecting internal failure. If the alarm remains after external causes are ruled out, remove the pump from service and document the findings clearly using the CCR method.

That is successful troubleshooting.