Cardinal Health Kangaroo OMNI

Alarm: System Error

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo OMNI

What This Guide Helps With

This guide helps troubleshoot Kangaroo OMNI System Error alarms that may stop pump operation, interrupt feeding or flushing, or indicate a device-level system fault.

A System Error should be treated differently from routine feeding set, cassette, or battery alarms because it may involve a broader pump operating issue.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Follow the On-Screen Restart Instruction

Confirm Whether the Alarm Returns

Check External and Simple Causes First

Check Power Source and Adapter Connection

Remove Accessories and Reload Only What Is Needed

Check for Repeatability

Stop When Resolved

If the Problem Persists

If the System Error returns after turning the pump off and back on, external and common causes have been ruled out.

Document the three-number code shown on the screen, press Power Off, remove the pump from service, and label it Out of Service.

A persistent System Error strongly indicates an internal system fault or device-level failure requiring repair, bench evaluation, or manufacturer support.

Do not continue repeated restarts in patient care areas. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a System Error on an active patient setup. Move the patient to a backup device first before continuing. A pump with an unresolved System Error should not be trusted for patient use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Kangaroo OMNI feeding pump displayed System Error during use. Pump stopped normal operation and required restart. Three-number code was present on screen."

Cause

What was observed during troubleshooting.

Example:
"System Error persisted after power cycle. AC power, external condition, cassette area, and feeding set setup were checked. Findings indicate likely internal device-level system fault."

Resolution

What action was taken.

Example:
"Recorded three-number System Error code, powered pump off, removed device from patient use, labeled Out of Service, and sent for repair or manufacturer support evaluation."

Helpful Details to Include (If Known)

Final Thought

A Kangaroo OMNI System Error should be approached as a device-reported fault, not a routine feeding setup alarm. Start with patient safety, confirm the exact alarm, perform the directed restart, check simple external causes, and document the displayed code if the error returns. Escalating a persistent System Error is the correct Clinical Engineering decision.

That is successful troubleshooting.