Cardinal Health Kangaroo OMNI

Alarm: Valve Error

On this page

Asset Type

Feeding Pump

Manufacturer

Cardinal Health

Model

Kangaroo OMNI

What This Guide Helps With

This guide helps troubleshoot Valve Error alarms on the Cardinal Health Kangaroo OMNI feeding pump. This alarm can stop feeding when the pump detects valve actuator resistance, blockage, contamination, misorientation, cassette seating problems, or feeding set issues.

A Valve Error should be treated as a delivery-interrupting condition because the pump may be unable to control flow correctly until the cassette and valve actuator area are inspected and corrected.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Translate the Alarm Into the Real-World Condition

Remove the Cassette

Inspect the Valve Actuator

Clean and Dry the Valve Actuator Area

Reload the Cassette Correctly

Replace the Feeding Set

Try a Different Kangaroo OMNI Pump

Stop When Resolved

If the Problem Persists

If the Valve Error alarm continues after the cassette is removed, the actuator is inspected, cleaned and dried, the cassette is reloaded, and the feeding set is replaced, external and common causes have been ruled out.

Persistent Valve Error alarms may indicate a device-side valve actuator or internal pump mechanism failure.

The pump should be removed from service, labeled Out of Service, and sent for repair or bench evaluation.

Do not keep clearing the alarm and returning the pump to clinical use.

Clinical Use Tip

Never troubleshoot a Valve Error on an active patient while feeding delivery is required. Move the patient to a backup feeding pump first, then evaluate the cassette, actuator, and feeding set. Avoid relying on a pump with unresolved valve actuator alarms.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Kangaroo OMNI feeding pump reported Valve Error alarm. Staff stated pump stopped feeding and alarm returned after attempting to continue."

Cause

What was observed during troubleshooting.

Example:
"Valve actuator area found with dried feeding solution and possible cassette binding. Alarm cleared after cassette removal, actuator cleaning/drying, and proper cassette reload."

Resolution

What action was taken.

Example:
"Removed cassette, inspected valve actuator for obstruction and orientation, cleaned and dried valve actuator area, reloaded cassette, and verified pump resumed operation without Valve Error alarm. Device returned to service after functional check."

Helpful Details to Include (If Known)

Final Thought

The Valve Error alarm should be treated as a delivery-interrupting condition. Start with patient safety, confirm the alarm, inspect the cassette and valve actuator area, clean and dry the actuator, and replace the feeding set if needed. If the alarm continues after these checks, remove the pump from service and escalate for repair. Accurate CCR documentation helps show what was tested, what was corrected, and why the final device status was appropriate.

That is successful troubleshooting.