Cardinal Health Kendall SCD 700 Series

Error: E1 System High Pressure Error

On this page

Asset Type

Sequential Compression Device (SCD)

Manufacturer

Cardinal Health

Model

Kendall SCD 700 Series

What This Guide Helps With

This guide addresses the E1 System High Pressure Error, which occurs when system pressure exceeds safe operating limits. This condition interrupts compression therapy and may be caused by airflow restrictions, improper garment use, or external factors affecting pressure regulation.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Inspect Tubing for Kinks or Obstructions

4. Check Garment Placement and Patient Interaction

5. Inspect Garments for Twisting or Compression

6. Check for Blocked or Damaged Connectors

7. Perform Manual Reset and Retest

If the Problem Persists

If the error continues after all external checks, external causes have been ruled out. The issue may indicate an internal pressure sensor, valve, or control failure.

Clinical Use Tip

Never troubleshoot compression devices while connected to a patient. Always ensure alternative DVT prevention is in place before removing therapy.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Device displaying E1 System High Pressure Error; compression therapy stopped during use."

Cause

What was observed during troubleshooting.

Example:
"Kinked tubing and patient positioning caused restricted airflow and elevated system pressure."

Resolution

What action was taken.

Example:
"Straightened tubing, repositioned patient, and reset device. Verified normal operation."

Helpful Details to Include (If Known)

Final Thought

High pressure errors are often caused by simple external restrictions. Always start with tubing and patient interaction before assuming device failure. Proper escalation and clear documentation ensure safe and effective resolution.

That is successful troubleshooting.