On this page
Asset Type
Sequential Compression Device (SCD)
Manufacturer
Model
What This Guide Helps With
Troubleshooting SCD Express blank display, keypad response, startup, power, alarm, or boot failure from external causes.
Step-by-Step Troubleshooting
Ensure Patient Safety First
Confirm the Kendall SCD Express is not actively providing compression therapy to a patient before troubleshooting.
If the patient requires continued DVT prevention therapy, have clinical staff move the patient to another working SCD pump first.
Expected outcome: Patient therapy is maintained and troubleshooting is performed safely.
Confirm the Reported Failure
Ask clinical staff what occurred during startup.
Check whether the issue is:
- Blank display
- Dim or flickering display
- Frozen screen
- Keypad buttons not responding
- Pump will not power on
- Unit powers on briefly then shuts down
- Startup alarm or abnormal tone
- Power indicator present with no display activity
Expected outcome: The problem is narrowed to power, display, keypad, or startup behavior.
Check the Wall Outlet
Plug another known-working device or outlet tester into the same receptacle.
Expected outcome: The outlet provides proper power.
If the outlet is dead or controlled by a wall switch, move the SCD to a verified outlet and retest. If this resolves the issue, stop.
Inspect the Power Cord and Plug
Check the AC power cord for cuts, bent prongs, loose strain relief, burn marks, or damage near the plug and pump connection.
Expected outcome: The cord is intact and seats securely.
If the cord is damaged, remove the device from service and replace the cord if facility policy allows. Retest after replacement. If this resolves the issue, stop.
Verify the Power Cord Is Fully Seated
Confirm the power cord is firmly connected to the rear of the SCD pump and the wall outlet.
A loose AC cord can cause intermittent startup, blank display, or unexpected shutdown.
Expected outcome: The unit receives stable AC power.
If reseating the cord restores normal startup, stop.
Test With a Known-Good Power Cord
Use a compatible known-good hospital-grade power cord if available.
Expected outcome: The SCD powers on normally with a known-good cord.
If the unit starts normally, document the failed cord and return the pump to service after safety checks.
Check for Visible Housing or Fluid Damage
Inspect the pump exterior, keypad surface, display lens, handle area, and rear power entry for cracks, fluid intrusion, sticky residue, or impact damage.
Expected outcome: No visible damage or contamination is present.
If fluid intrusion, cracked housing, burning smell, or unusual heat is found, remove the device from service and escalate for repair. Do not continue power cycling.
Perform a Controlled Power Cycle
Turn the unit off, disconnect AC power, wait briefly, reconnect AC power, and attempt startup again.
Expected outcome: The unit completes startup and the display/keypad responds normally.
If startup returns to normal and no alarms remain, verify operation with sleeves before returning to service.
Check Keypad Response
With the device powered on, press each user-accessible control normally. Do not press excessively hard or use tools.
Expected outcome: Buttons respond consistently without sticking or requiring abnormal force.
If only one keypad area is unresponsive, inspect for residue, cracked overlay, or depressed buttons. If the keypad remains unresponsive, remove the device from service.
Check Display Visibility
View the display from normal operating angles and under normal room lighting.
Expected outcome: The screen is readable and stable.
If the unit appears to operate but the display is unreadable, dim, flickering, or missing segments, remove it from service because therapy settings and alarms may not be reliably confirmed.
Check for Startup Alarm Behavior
Listen for abnormal tones, repeated beeps, or startup alarms.
Expected outcome: The device starts without unresolved alarm conditions.
If the device alarms immediately at startup, note the displayed message if visible. If the message cannot be cleared through normal user controls, remove the device from service.
Inspect Tubing and Sleeve Connections
Connect known-good SCD tubing and sleeves if the unit powers on but will not complete startup or appears to stall during readiness checks.
Expected outcome: The device recognizes connected accessories and proceeds normally.
If a bad tube or sleeve caused the startup concern, replace the accessory and stop.
Check for Mechanical Obstruction or Pump Noise
During startup, listen for unusual clicking, grinding, repeated motor attempts, or no pump activity at all.
Expected outcome: Startup sounds are normal and brief.
If the display/keypad issue is paired with abnormal internal noise, remove the device from service and send for bench evaluation.
Verify Final Operation Before Return to Service
After any successful correction, power the unit on, confirm the display is readable, keypad responds, alarms are clear, and the pump runs with known-good tubing and sleeves.
Expected outcome: The Kendall SCD Express starts normally and is safe to return to clinical use.
If the Problem Persists
If the Kendall SCD Express still has a blank display, frozen screen, unresponsive keypad, intermittent startup, abnormal alarm, unusual heat, burning smell, or repeated shutdown after external checks, common external causes have been ruled out.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
At that point, the issue may involve the display assembly, keypad overlay, power entry, internal power supply, control board, or startup circuitry. Knowing when to stop is proper troubleshooting.
Clinical Use Tip
Do not troubleshoot a Kendall SCD Express while it is connected to an active patient. Move the patient to another working compression pump first, especially if compression therapy is ordered for DVT prevention.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported the Kendall SCD Express would not start and the display remained blank when plugged in."
Cause
What was observed during troubleshooting.
Example:
"Verified wall power and found the AC power cord connection loose at the rear of the SCD pump."
Resolution
What action was taken.
Example:
"Reseated the power cord, completed startup test with known-good sleeves, confirmed display and keypad operation, and returned the unit to service."
Helpful Details to Include (If Known)
- Outlet tested
- Power cord inspected or swapped
- Display blank, dim, frozen, or flickering
- Keypad response confirmed or failed
- Startup tones or alarms observed
- Tubing and sleeves tested
- Indicator lights present or absent
- Any fluid damage, cracks, heat, smell, or impact damage
- Final device status: returned to service or removed from service
Final Thought
For display, keypad, and startup failures, begin with patient safety, then verify power, cord condition, startup behavior, accessories, and visible damage before assuming internal failure. Clear documentation helps show what was checked, what was ruled out, and why the device was either returned to service or escalated for repair.
That is successful troubleshooting.